Apera PH700 Instruction Manual Download Page 12

 

- 10 - 

5.5.1 According to the pH isothermal measurement principle, the closer the test sample’s temperature is to 
the calibration solution’s, the higher the accuracy of the measurement. This principle is recommended to 
follow when conducting tests for the best result. Example: If users were to test samples at 150˚F, we 
recommend warming up the calibration solutions to the same temperature before performing calibrations in 
order to get the most accurate readings. 

5.5.2 The instrument has a function to return to factory default setting, which can be set up in P5 (refer to 
section 5.6). Returning to factory default setting is to restore the meter to theoretical value (zero potential pH 
is 7.00, slope is 100%), and set all the parameters to default settings. When the meter’s calibration or 
measurement is performing abnormally, users can use this function to let the meter return to factory default 
mode, and then conduct calibration and test again. Please note that this function is irreversible once used. 

5.5.3 Self-Diagnosis 

In the process of calibration and measurement, the meter has a self-diagnosis function. Please see table 5. 

Table – 5 Self-diagnosis Information 

Display 

Icons 

Contents 

How to fix 

 

Wrong pH buffer solution or the buffer 
solution is out of range. 

1. 

Check whether pH buffer solution is correct. 

2. 

Check  whether  the  meter  connects  the 
electrode properly. 

3. 

Check whether the electrode is damaged. 

4. 

Make  sure  there  is  no  bubble  in  the 
electrode’s bulb. 

 

pH  electrode  zero  electric  potential 
out of range (<-60mV or >60mV) 

 

1. 

Make sure there is no bubble in the 
electrode’s bulb. 

2. 

Check whether the electrode is damaged. 

3. 

Check if pH buffer solutions are correct.   

4. 

The pH Electrode is aged. Replace a new 
one. 

 

pH electrode slope is out of range 
(<85% or >110%) 

 

Press          key  when  reading  is  not 
stable during calibration. 

Press          key

 

when          icon appears and stays 

on the screen. 

 

During calibration, the measuring value 
is not stable for ≥3min. 

1. 

Make sure there is no bubble in the 
electrode’s bulb. 

2. 

Check whether the electrode is damaged. 

3. 

The pH Electrode is aged. Replace a new 
one. 

 

 

* If you find any air bubble in the glass bulb of the pH sensor, simply shake the probe for a few times 
to remove it. The existence of an air bubble in the glass bulb will significantly decrease the stableness 
of measurement.   

 

 

 

 

 

Summary of Contents for PH700

Page 1: ...PH700 Benchtop pH Meter Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de Manual Version 3 2 ...

Page 2: ... 4 5 Data storage recall and deletion 4 4 6 Max Min Function 5 5 pH Measurement 5 5 1 Preparation 5 5 2 pH Electrode Information 5 5 3 Information regarding Calibration 6 5 4 pH Meter Calibration 8 5 5 Measuring pH 9 5 6 Parameter Setting 11 5 7 pH Electrode s Maintenance 12 6 mV measurement 13 6 1 ORP measurement 13 6 2 Notes on ORP measurement 13 7 Recommended pH Electrodes for Specific Applicat...

Page 3: ...essing technology intelligently improves the meter s response time and measuring accuracy with stable reading display mode available The kit comes with a 3 in 1 combination pH Electrode measuring pH and temperature simultaneously IP54 dustproof and spillproof Connectors are protected by silicone seal caps effectively ensuring the reliability and service life of the meter under harsh environment De...

Page 4: ...on icons and numberings 3 Measurement unit 4 Temperature and reminder icons 5 The pH unit displayed during calibration 6 The pH value displayed during calibration numberings for storage and recall and reminder icons 7 Icons for data storage and recall M store measuring data RM Recall measuring data 8 Self Diagnosis icons and numberings 9 Temperature compensation mode icons ATC Auto Temperature Com...

Page 5: ...ion and go back to measurement mode Short Press In calibration mode press to conduct calibration In parameter setting press to confirm selection Short Press Store measuring data Long Press Recall stored measuring data Short Press or Long Press In manual temperature compensation mode short press to change temperature long press to change quickly In parameter setting mode short press to change param...

Page 6: ...d deletion Storage When the measurement is stable press key the meter displays M icon and storage numberings on LCD and stores the measuring data In Figure 4 the meter stores the first group of the measuring data Press again to keep storing Recall stored data In the measurement mode long press key to recall the last stored Measuring data The meter displays RM icon and storage numberings In Figure ...

Page 7: ...the electrode holder The electrode holder is composed of a metal stand base a supporting pole and an electrode clip First install the metal stand base to the meter screw it on Please note that the metal base can be installed on either side of the meter Then install the supporting bar on the metal stand base and then install the electrode clip on the supporting bar Connect the power adapter Plug th...

Page 8: ...f excess water Gently stir the electrode for a few seconds after it s dipped into the test solution and then let it stand still Wait for the stable measurement a smiley face appears and stays and then record the readings When test is finished place the electrode in the storage bottle and tighten the cap in order to keep the sensitivity of the pH sensor 5 3 Information regarding Calibration pH buff...

Page 9: ...ondition of electrodes and the requirement of the accuracy For High Accuracy meaustements 0 02pH the meter should be calibrated before test every time For ordinary accuracy measurements 0 1pH once calibrated the meter can be used for about a week or longer In the following cases the meter must be re calibrated a The electrode hasn t been used for a long time or a new electrode is connected b After...

Page 10: ...ed water dry it and dip it into pH 4 00 buffer solution Stir the solution gently and let stand still in the buffer solution until a stable reading is reached The meter s display will show the recognition process of calibration buffer solution at the lower right corner When the meter recognizes 4 00 pH stable icon displays on LCD Press key to calibrate the meter End icon and electrode slope of acid...

Page 11: ...e Contact us at 1 614 285 3080 for more details When testing purified water like spring water or drinking water it will take longer for the readings to get stabilized typically 3 5 minutes because there is very few ions left to be detected by the sensor in those purified water 2 Purified water such as distilled water and deionized water are recommended for only rinsing the probe for the best resul...

Page 12: ...ation and measurement the meter has a self diagnosis function Please see table 5 Table 5 Self diagnosis Information Display Icons Contents How to fix Wrong pH buffer solution or the buffer solution is out of range 1 Check whether pH buffer solution is correct 2 Check whether the meter connects the electrode properly 3 Check whether the electrode is damaged 4 Make sure there is no bubble in the ele...

Page 13: ... Setting P1 Select pH buffer series USA NIS 1 Long press to enter P1 2 Press USA flashes press to select USA or NIS press to confirm USA USA Series NIS NIST Series 3 Press to enter P2 or press to return to measurement mode P2 Select resolution 0 01 0 1 1 Press 0 01 flashes press to select 0 01 or 0 1 press to confirm 2 Press to enter P3 or press to return to measurement mode P3 Select temperature ...

Page 14: ...recommend 10 15 times of use before replacing the pH buffers 5 7 3 Protect glass bulb The sensitive glass bulb at the front of the combination electrode should not come in contact with hard surfaces Scratches or cracks on the electrode will cause inaccurate readings Before and after each measurement the electrode should be rinsed with distilled water or purified water If a sample sticks to the ele...

Page 15: ...227 222 215 209 205 201 6 2 2 Clean and activate ORP electrode After the electrode has been used over a long period of time the platinum surface will get polluted which causes inaccurate measurement and slow response Please refer to the following methods to clean and activate ORP electrode a For inorganic pollutant submerge the electrode in 0 1mol L dilute hydrochloric acid for 30 minutes then was...

Page 16: ...e warrant this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge at option of APERA INSTRUMENTS Europe GmbH any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS Europe GmbH for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does not cover any damages due to Transpo...

Reviews: