background image

 

- 7 - 

6.8

 For the self-diagnosis information, please refer to the table below: 

Symbol 

Self-Diagnosis information 

Potential problems and how to fix 

 

 

 

 

The pH calibration solution cannot be 

recognized by the meter. 

1. 

Make sure the probe is fully immersed in the calibration 
solution. 

2. 

Check if calibration solution is expired or polluted. 

3. 

1st point of pH calibration must be pH 7.00 or 6.86. See 6.6. 

4. 

Please check whether pH probe is damaged or broken. If 
so, please replace with a new one. 

5. 

The glass bulb or junction is severely contaminated. Please 
use a soft brush with soap water to clean it thoroughly. Then 
soak it in 3M KCL 3-5 hours before performing calibration 
again. 

 

 

 

      is pressed before measurement   

is fully stabilized. 

Wait for          to stay on screen before pressing 

 

During calibration, readings being unstable for 

over 3 minutes. 

1. 

Please check whether pH probe is damaged or broken. If so, 
please replace with a new one. 

2. 

The glass bulb or junction is severely contaminated. Please 
use a soft brush with soap water to clean it thoroughly. Then 
soak it in 3M KCL overnight before performing calibration 
again. 

3. 

The probe is aged (used for over a year and has a much 
slower response). A replacement is needed. 

 

pH probe zero electric potential out of range 

(<-60mV or >60mV) 

1. 

Check whether pH buffer solutions comply with the USA or 
NIST standard. 

2. 

Check whether pH buffers are expired or contaminated. 

3. 

Please check whether pH probe is damaged or broken. If so, 
please replace with a new one. 

4. 

The probe is aged (used for over a year and has a much 
slower response). A replacement is needed. 

5. 

The probe is invalidated (Er4/Er5 repetitively appears, and 
problems 1, 2, 3 are excluded). Probe replacement is 
needed. 

 

pH probe slope out of range 

(<85% or >110%) 

 

The calibration reminder is triggered. I

t’s time to 

perform a new pH calibration. 

Perform pH calibration or cancel calibration reminder in ZenTest App 

settings. 

 

 

pH Measurement  

7.1   

PH60S-Z Spear pH tester is designed for direct spear test of solid samplings such as cheese, fruit, sushi rice, meat, 

dough. It also works well in general water solutions.   

7.2

    When performing spear tests in solid samples, pay attention to insert probe evenly, and be careful to prevent probe 

from damage. If the medium is too hard, please bore a small hole with a clean knife before inserting the spear probe. 

 

For food pH measurement (e.g. sushi, cheese, meat, fruit, etc.), it should always be a sampling test. That means 

test samples should no longer be edible. 

Summary of Contents for ZenTest PH60S-Z

Page 1: ...1 ZenTest PH60S Z Smart Spear pH Tester pH Temp ORP User Manual APERA INSTRUMENTS LLC aperainst com V1 5...

Page 2: ...E 5 PH CALIBRATION 5 PH MEASUREMENT 7 ORP MEASUREMENT 8 PROBE CLEANING 8 PROBE STORAGE 8 PARAMETER SETTING 9 TECHNICAL SPECIFICATIONS 10 ICONS AND FUNCTIONS 10 TROUBLESHOOTING GUIDE 11 PROBE REPLACEME...

Page 3: ...erate the tester without connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LED display 1 Basic Mode digital display cal...

Page 4: ...to turn on off backlight 1 In measurement mode long press to turn on off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 3 In parameter setting short press...

Page 5: ...water RO water or tap water is the alternative then shake off excess water see graph on the right 5 3 Perform at least a 2 point calibration For pH calibration tutorial refer to Section 6 5 4 When you...

Page 6: ...librate 2nd or 3rd point Otherwise you will need to restart the calibration process with 7 00 pH first 2 The pH 4 00 and 7 00 buffer solutions poured into the calibration vials can be used for up to 1...

Page 7: ...d for over a year and has a much slower response A replacement is needed pH probe zero electric potential out of range 60mV or 60mV 1 Check whether pH buffer solutions comply with the USA or NIST stan...

Page 8: ...inal probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 15 for how to replace a probe 3 Rinse the probe in distilled water and dry...

Page 9: ...if 15 minutes is set the meter will automatically shut down after 15 minutes if no operation when Off is set the auto power off function will be turned off Long press to manually shut down the meter...

Page 10: ...unctions Calibrated points Self Diagnosis Symbol Er1 Er2 Er3 Er4 Er5 Er6 Stable reading indicator Waterproof Rating IP67 floats on water Reading Lock HOLD Power DC3V AAA batteries 4 Bluetooth Signal B...

Page 11: ...ble damage replace the probe Instrument defect Contact us for warranty fulfillment Probe is not fully submerged in the solution Make sure the probe is immersed into solution at least 1 inch Air bubble...

Page 12: ...ged product attributable to responsibility of APERA INSTRUMENTS LLC for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to Acciden...

Reviews: