6.4

 Insert the probe into pH 7.00 buffer solution, make a quick stir, and hold still. When the reading is stabilized (          stays 

on  screen),  short  press            to  start  1

st

  point  calibration.  After  calibration  is  completed,  the  tester  will  return  to 

measurement mode. Icon          will appear at the bottom left of the screen, indicating a successful 1

st

 point pH calibration. 

6.5

 To calibrate 2nd point, use 4.00 pH buffer and repeat Step 6.3 to 6.4 (Do NOT turn off the tester after you finish pH 7 

calibration).            will display next to          , indicating a successful 2-point pH calibration (low and middle points).

   

 

6.6

 If necessary (target pH>8.00), calibrate 3rd point using 10.01 standard pH buffer and repeat Step 6.3 to 6.4,          will 

show up next to            and          , indicating a successful 3-point calibration (high, low, and middle points). 

 

6.7

  Notes about Calibration 

1)  The 1st point calibration must be 7.00 pH. Perform the 2nd and 3rd point calibrations (4.00, 10.01, 1.68, or 12.45) 

immediately after the 1st point calibration is finished. Do NOT turn off the meter before you calibrate 2nd or 3rd 

point. Otherwise, you will need to restart the calibration process with 7.00 pH first. 

2)  The pH 4.00 and 7.00 buffer solutions poured into the calibration vials can be used for up to 10 times as long as 

they are not contaminated and the bottles are capped when not in use. pH 10.01 can only be used for up to 5 

times as it will lose its accuracy much faster. After that, replace the buffer solutions in the calibration vials with 

new  ones  to  keep  the  accuracy.  Keeping  the  freshness  and  cleanliness  of  calibration  buffers  is  essential  for 

accurate pH measurement. 

3)  The tester can perform 1 to 3 points of automatic calibration and can recognize 5 types of pH standard solutions. 

For details, please refer to the following table: 

 

Calibration 

USA Series 

NIST Series 

Indication icon 

Recommended 

1-pt 

7.00 pH 

6.86 pH 

 

 

Accuracy ≥ 0.1 pH 

2-pt 

Option A 

1st pt: 7.00 pH 

2nd pt: 4.00 or 1.68 pH 

Option A 

1st pt: 6.86 pH 

2nd pt: 4.01 or 1.68 pH 

 

 

Range < 7.00 pH 

Option B 

1st pt: 7.00 pH 

2nd pt: 10.01 or 12.45 pH 

Option B 

1st pt: 6.86 pH 

2nd pt: 9.18 or 12.45 pH 

 

 

Range >7.00 pH 

3-pt 

1st pt: 7.00 pH 

2nd pt: 4.00 or 1.68 pH 

3rd pt: 10.01 or 12.45 pH 

1st pt: 6.86 pH 

2nd pt: 4.01 or 1.68 pH 

3rd pt: 9.18 pH or 12.45 pH 

 

 

Range: 0 to 14.00 pH 

 

 

 

 

 

 

Summary of Contents for ZenTest PH60S-Z

Page 1: ...1 ZenTest PH60S Z Smart Spear pH Tester pH Temp ORP User Manual APERA INSTRUMENTS LLC aperainst com V1 5...

Page 2: ...E 5 PH CALIBRATION 5 PH MEASUREMENT 7 ORP MEASUREMENT 8 PROBE CLEANING 8 PROBE STORAGE 8 PARAMETER SETTING 9 TECHNICAL SPECIFICATIONS 10 ICONS AND FUNCTIONS 10 TROUBLESHOOTING GUIDE 11 PROBE REPLACEME...

Page 3: ...erate the tester without connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LED display 1 Basic Mode digital display cal...

Page 4: ...to turn on off backlight 1 In measurement mode long press to turn on off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 3 In parameter setting short press...

Page 5: ...water RO water or tap water is the alternative then shake off excess water see graph on the right 5 3 Perform at least a 2 point calibration For pH calibration tutorial refer to Section 6 5 4 When you...

Page 6: ...librate 2nd or 3rd point Otherwise you will need to restart the calibration process with 7 00 pH first 2 The pH 4 00 and 7 00 buffer solutions poured into the calibration vials can be used for up to 1...

Page 7: ...d for over a year and has a much slower response A replacement is needed pH probe zero electric potential out of range 60mV or 60mV 1 Check whether pH buffer solutions comply with the USA or NIST stan...

Page 8: ...inal probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 15 for how to replace a probe 3 Rinse the probe in distilled water and dry...

Page 9: ...if 15 minutes is set the meter will automatically shut down after 15 minutes if no operation when Off is set the auto power off function will be turned off Long press to manually shut down the meter...

Page 10: ...unctions Calibrated points Self Diagnosis Symbol Er1 Er2 Er3 Er4 Er5 Er6 Stable reading indicator Waterproof Rating IP67 floats on water Reading Lock HOLD Power DC3V AAA batteries 4 Bluetooth Signal B...

Page 11: ...ble damage replace the probe Instrument defect Contact us for warranty fulfillment Probe is not fully submerged in the solution Make sure the probe is immersed into solution at least 1 inch Air bubble...

Page 12: ...ged product attributable to responsibility of APERA INSTRUMENTS LLC for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to Acciden...

Reviews: