Apera Instruments ZenTest PH60-Z User Manual Download Page 9

 

- 9 - 

 

Probe St orage  

1)  Under regular usage (daily or weekly use), make sure the probe cap is wet, and tightly close the cap with the O-ring. 

2)  For long-term storage (you are not going to use the product for a while),    add 3M KCL soaking solution to the Fill 

line in the probe cap and store the probe in it. Close on the probe cap tightly with the O-ring. 

3)  If you find white crystals inside or outside the probe cap, it is perfectly normal. It is the 3M KCL soaking solution that 

crystalizes over time by its nature. Just rinse them off and add in new soaking solution. This chemical is not 

poisonous nor dangerous, and the probe’s performance will not be affected at all. 

4)  NEVER store the probe in pure water like tap, RO, distilled, or deionized water as they could damage the pH probe. 

If this happens, immediately soak the pH probe Apera 3M KCL soaking solution overnight, then re-calibrate it before 

using. Pure water is only for rinsing the probe. 

 

Paramet er Setting  

 

Table of Settings 

 

Parameter Setting 

1) 

When the meter is turned off, long press           

to enter parameter setting →short press            to switch P01-P02… 

→P8. Short Press          , parameter flashes → short press            to adjust parameter → short press            to confirm 

→Short press          to exit parameter setting and go back to measurement mode. 

2) 

Auto. Lock (P02)

 

– Users can set the auto lock time from 5 to 20 seconds. For example, if 10 seconds is set, when 

the measured value is stable for more than 10 seconds, the measured value will be automatically locked, and the 

HOLD icon will be displayed. Short press          to release the lock. When the setting is "Off" , the Auto. lock function 

is turned off, that is, the measured value can only be manually locked. Short press          to lock or unlock the 

measured value. The HOLD icon will be displayed when reading is locked. 

3) 

Auto. Backlight (P03)

 

  Users can set the automatic backlight time for 1 to 8 minutes. For example, if 3 minutes 

is set, the backlight will turn off automatically after 3 minutes; when the 

“Off” is set, the auto. backlight function will 

be turned off, and short press          to manually turn the backlight on or off. 

Symbol 

Parameter Setting Contents 

Content 

Factory Default 

P1 

Temperature Unit 

˚C – ˚F 

˚F 

P2 

Select automatic lock 

5-20 seconds 

– Off 

Off 

P3 

Automatic Backlight Off 

1-8 minutes

 

– Off 

P4 

Automatic Power Off 

10-20 minutes 

– Off 

10 

P5 

pH Buffer Series Selection 

USA 

– NIST 

USA 

P6 

pH Resolution 

0.1 

– 0.01 

0.01 

P7 

pH Calibration Reminder 

H-hours D-Days (set up in ZenTest App) 

P8 

pH back to factory default 

No

 

– Yes 

No 

Summary of Contents for ZenTest PH60-Z

Page 1: ...1 ZenTest PH60 Z Smart pH Tester Kit pH Temp ORP User Manual APERA INSTRUMENTS LLC aperainst com V1 5...

Page 2: ...2 ATTENTION Water droplets are added during production to maintain the moisture of the probe This is normal practice and should not be attributed to used product Never use the product when it s freezi...

Page 3: ...erate the tester without connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LED display 1 Basic Mode digital display cal...

Page 4: ...t press to turn on off backlight 1 In measurement mode long press to turn on off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 3 In parameter setting sho...

Page 5: ...ARDS Preparation before Use 5 1 Pull out the battery insulation slip and take off the probe cap 5 2 Rinse off the probe in pure water preferably distilled or deionized water RO water or tap water is t...

Page 6: ...0 01 standard pH buffer and repeat Step 6 3 to 6 4 will show up next to and indicating a successful 3 point calibration high low and middle points 6 7 Notes about Calibration 1 The 1st point calibrati...

Page 7: ...n soak it in 3M KCL 3 5 hours before performing calibration again is pressed before measurement is fully stabilized Wait for to stay on screen before pressing During calibration readings being unstabl...

Page 8: ...sured in mV It s also called redox ORP is a measure of the cleanliness of water its ability to break down contaminants A separate ORP probe ORP60 DA needs to be installed to be able to measure ORP 2 P...

Page 9: ...ort press to adjust parameter short press to confirm Short press to exit parameter setting and go back to measurement mode 2 Auto Lock P02 Users can set the auto lock time from 5 to 20 seconds For exa...

Page 10: ...Factory Default P08 Select Yes to recover instrument calibration to theoretical value This function can be used when instrument does not work well in calibration or measurement Calibrate and measure a...

Page 11: ...ble damage replace the probe Instrument defect Contact us for warranty fulfillment Probe is not fully submerged in the solution Make sure the probe is immersed into solution at least 1 inch Air bubble...

Page 12: ...ged product attributable to responsibility of APERA INSTRUMENTS LLC for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to Acciden...

Reviews: