Apera Instruments ZenTest PH60-Z User Manual Download Page 5

 

- 5 - 

 

  Batt ery Replacement  

Please install batteries according to the following steps. 

*

Please note the correct direction of battery installation:   

The Positive Side 

(“+”) OF EVERY SINGLE Battery MUST FACE UP

.   

(WRONG INSTALLATION OF BATTERIES WILL CAUSE DAMAGE TO THE TESTER AND POTENTIAL HAZARDS!)

 

 

 

 

 

 

 

 

 

 

 

 

Preparation bef ore Use  

5.1

    Pull out the battery insulation slip, and take off the probe  cap. 

 

5.2

    Rinse off the probe in pure water (preferably distilled or deionized water. RO water or tap water is the alternative) , 

then shake off excess water.

 

 

 

 

 

 

 

 

 

 

 

 

5.3

    Perform calibration. For pH calibration tutorial, refer to Section 6. 

   

5.4

    If the tester 

hasn’t been used for a long time (over 1 month) , please soak the probe in the 3M KCL 

soaking solution for 15 minutes, then calibrate it before test. 

 

Summary of Contents for ZenTest PH60-Z

Page 1: ...1 ZenTest PH60 Z Smart pH Tester Kit pH Temp ORP User Manual APERA INSTRUMENTS LLC aperainst com V1 5...

Page 2: ...2 ATTENTION Water droplets are added during production to maintain the moisture of the probe This is normal practice and should not be attributed to used product Never use the product when it s freezi...

Page 3: ...erate the tester without connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LED display 1 Basic Mode digital display cal...

Page 4: ...t press to turn on off backlight 1 In measurement mode long press to turn on off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 3 In parameter setting sho...

Page 5: ...ARDS Preparation before Use 5 1 Pull out the battery insulation slip and take off the probe cap 5 2 Rinse off the probe in pure water preferably distilled or deionized water RO water or tap water is t...

Page 6: ...0 01 standard pH buffer and repeat Step 6 3 to 6 4 will show up next to and indicating a successful 3 point calibration high low and middle points 6 7 Notes about Calibration 1 The 1st point calibrati...

Page 7: ...n soak it in 3M KCL 3 5 hours before performing calibration again is pressed before measurement is fully stabilized Wait for to stay on screen before pressing During calibration readings being unstabl...

Page 8: ...sured in mV It s also called redox ORP is a measure of the cleanliness of water its ability to break down contaminants A separate ORP probe ORP60 DA needs to be installed to be able to measure ORP 2 P...

Page 9: ...ort press to adjust parameter short press to confirm Short press to exit parameter setting and go back to measurement mode 2 Auto Lock P02 Users can set the auto lock time from 5 to 20 seconds For exa...

Page 10: ...Factory Default P08 Select Yes to recover instrument calibration to theoretical value This function can be used when instrument does not work well in calibration or measurement Calibrate and measure a...

Page 11: ...ble damage replace the probe Instrument defect Contact us for warranty fulfillment Probe is not fully submerged in the solution Make sure the probe is immersed into solution at least 1 inch Air bubble...

Page 12: ...ged product attributable to responsibility of APERA INSTRUMENTS LLC for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to Acciden...

Reviews: