will cause big error to turbidity measurement. So, the vial should be left stand still for 2 to 5 minutes 

to  eliminate  potential  air  bubbles  before  measuring.  We  recommend  using  the  same  mixing 

procedure and the same waiting time for degassing before comparing measurements.

 

 

d) 

Other Requirements 

 

On the premise of ensuring evenly distributed samples, sample solution should be measured 

immediately to prevent temperature changing and precipitates from affecting measurements. 

 

Avoid sample dilution for measurement as much as possible. 

 

Avoid operating under direct sunlight. 

 

Do not pour solution into the sample vial holder. Sample vials must be used for measurement. 

 

Please do not wash the sample vial holder as it may damage the optical structure. 

 
 

WARRANTY 

We warrant this instrument to be free from defects in material and workmanship and agree to repair or 

replace free of charge, at the option of APERA INSTRUMENTS, LLC, any malfunctioned or damaged 

product attributable to the responsibility of APERA INSTRUMENTS, LLC for a period of TWO YEARS for 

the instrument from the delivery. 

This limited warranty does NOT cover any issues due to: 

 

Accidental damage 

 

Improper use 

 

Normal wear and tear 

 

Transportation 

 

Storage 

 

Failure to follow the product instructions 

 

Unauthorized maintenance, modifications, combination or use with any products, materials, 

processes, systems or other matter 

 

Unauthorized repair 

 

External causes such as accidents, abuse, or other actions or events beyond our reasonable 

control. 

 

 

Summary of Contents for TN400

Page 1: ...1 TN400 Portable Turbidity Meter Instruction Manual APERA INSTRUMENTS LLC aperainst com...

Page 2: ...Display Mode 6 3 4 Keypad 7 3 5 Power Supply 7 4 Meter Calibration 8 4 1 Preparation for Calibration 8 4 2 Calibration Procedure Take 0 NTU and 20 NTU as an example 9 4 3 Notes for Calibration 10 5 T...

Page 3: ...characteristics Smart functions such as automatic calibration operating instructions automatic power off low voltage display etc Large TFT color screen with blue background in measuring mode green bac...

Page 4: ...s AMCO Polymer or Formazin Solution 0 20 100 400 and 800 NTU FNU Detector Silicon photovoltaic Measurement Mode Normal measurement and average measurement mode Display TFT color screen Sample Vials 25...

Page 5: ...vial holder Close the cover when measuring Dust proof plug Take off the plug when measuring Housing Sample vial holder Display Calibration vials or sample vials Keypad Positioning mark Align the mark...

Page 6: ...0 NTU Carrying case TN400 Turbidimeter Microfiber cloth Screwdriver to remove battery cover Silicone oil 10 ml AA Alkaline battery 4 Sample vial 6 Instruction manual Diagram 2 Measurement mode Measure...

Page 7: ...erform single shot measurement long press the key to perform continuous measurement In calibration mode press the key to confirm the calibration In calibration mode press the key to select standard so...

Page 8: ...your supplier to replace See the detail in Section 7 Replacement Parts b Replace zero turbidity solution Replace with 0 0 NTU calibration solution open the vial cap pour out the original solution add...

Page 9: ...the vial to make the solution even For polymer calibration solutions that have not been used for a long time slowly flip the vial twice and let it stand for 2 minutes Be careful not to shake the solu...

Page 10: ...00 NTU calibration vial in the sample vial holder and press to calibrate After calibration is done press to confirm repeat the steps to calibrate 400 NTU and 800 NTU After calibration is done press to...

Page 11: ...ation solution If the wrong solution is selected for calibration the measurement will be completely wrong To fix this problem simply re calibrate the meter selecting the correct calibration solutions...

Page 12: ...s of continuous measurement displaying 1 data every 8 seconds and will display the average value at the end see diagram 6 The average measurement mode can be used for observing the stabilizing process...

Page 13: ...d for measurement Please do not wash the sample vial holder as it may damage the optical structure 6 WARRANTY We warrant this instrument to be free from defects in material and workmanship and agree t...

Page 14: ...L 1 bottle T200 1 20 100 400 800 NTU AMCO polymer Compatible with TN400 TN480 4 bottles T500 3 Sample vials with lid 25 60mm 1 6 6 pcs TN400 S3 Silicone oil 10mL 1 bottle APERA INSTRUMENTS LLC Address...

Reviews: