7.3 Parameter setting notes 

a) 

TDS Factor (P1): 
Users can adjust TDS Factor by experimental data or experience. The 

following chart lists some commonly used TDS Factors for reference. 

 

Conductivity and TDS Factor 

Conductivity of the solution 

TDS Factor 

0~100 

μS/cm 

0.60 

100~1000 

μS/cm 

0.71 

1~10 mS/cm 

0.81 

10~100 mS/cm 

0.94 

 

b) 

P2-Select temperature unit (°C-°F): 

In P1 mode, press 

 to enter P2, press  

, °C or °F flickering, press  

 to 

choose °C-°F, press 

 to confirm parameter selection. Long press 

 to return 

to measurement mode. 

c) 

Restore to factory default: 
Select Yes to restore the calibration to the theoretical values and parameter 

setting to original values. When meter’s calibration or measurement performs 

abnormally, this function can be adopted so the meter goes back to factory 

default setting and then users can conduct calibration or take measurements 

again. 

8.  Technical Specifications 

TDS 

Range 

0

–100.0 ppm, 0–1000 ppm, 0–10.00 ppt 

Resolution 

0.1/1 ppm, 0.01 ppt 

Accuracy 

±1% F.S 

Calibration points 

1-2 points 

Automatic Temp. 
Compensation 

0-

50˚C 

Temp. 

Range 

0-

50˚C 

Resolution 

0.1˚C 

Accuracy 

±0.5˚C 