Using solid electrolyte as reference solution, it does not contanminate
testing samples, suitable for testing of samples that are rich in fat and
proteins.
n
Special notice
:
a) Electrode needs to be soaked in Polymer solution when not in use
,
do
not store the electrode in dry environment. Polymer solution (50ml) is provided
with the meter.
b) To prevent damaging the electrode, please bore a hole in harder
samples and then insert the electrode into the hole for testing.
c) In order to prevent rusting, please do not scratch the stainless steel
casing of the electrode with sharp object or sand paper.
2 Technical Specifications
2.1 Meter’s Technical Specs
Technical Specs
pH
Range
-2.00 to 19.99 pH
Resolution
0.01/0.1 pH
Accuracy
±0.01 pH ±1 digit
Temperature
Compensation
0 to 100 ˚C (Auto or Manual)
mV
Range
-1999 mV - 0 - 1999 mV
Resolution
- 200 mV - 0 - 200 mV: 0.1 mV; Remaining: 1 mV
Accuracy
±0.1% F.S ±1 digit
Temp.
Range
0 to 100˚C
Resolution
0.1˚C
Accuracy
±0.5˚C ±1 digit
2.2 Others
Data
Storage
500 groups