Apera Instruments SX811-BS Instruction Manual Download Page 17

 

- 14 - 

 

P1.4. – Set pH measurement of pure water (Off – On) 

1.  Press        key, 

Off

 flashes, then press

   

    key, 

On

 flashes, when   

parameter flashes, press        key to confirm. 

Off

–turn off temperature 

compensation, 

On

 – turn on temperature compensation. 

2.  After confirm parameter, press   

     

key to enter in P1.5 mode, or 

press            key to return to the measurement mode. 

 

P1.5. – Set pH measurement of pure water with ammonia (Off – On) 

1. Press          key, 

Off

 flashes, then press

   

      key, 

On

 flashes, when   

parameter flashes, press          key to confirm. 

Off

–turn off temperature   

compensation, 

On

–turn on temperature compensation. 

2. After confirm parameter, press   

     

key to enter in P1.6 mode, press 

          key to return to the measurement mode. 

 

P1.6. – Restore to factory default setting (NO – Yes) 

1.  Press       key, 

No

 flashes, then press       key, 

Yes

 flashes. Press 

key to confirm, the meter returns to the measurement mode. 

No 

– Not   

restore to factory default setting, 

Yes

 – restore to factory default setting. 

2.  When do not select 

Yes

, press          key to return to the measurement 

mode. 

 

 

 

6.4  Submenu of basic parameter setup (press        key or        key to switch) 

 

P4.1. – Adjust timing storage time 

1. In mode P4.0, press 

     

  key to enter in mode P4.1, refer to the left 

diagram: “ 

00: 

”: hours (0-99), “ 

:00 

”: minutes (0-59). 

2. Press          key, “ 

:00 

” flashes, then press   

   

  key, “ 

00: 

” flashes. 

When the number flashes, press   

   

  key and

       

  key to adjust time 

and press   

   

  key to confirm. 

3. After confirm the parameter, press 

     

  key to enter in P4. 2 mode or 

press 

     

  key to return to the measurement mode. 

 

 

P4.2. – Select temperature unit (°C—°F). 

1. Press 

       

key, °C flashes, then press   

   

 key, °F flashes. When the 

parameter flashes, press 

       

key to confirm. 

2. After confirm the parameter, press

     

    key to enter in P4.3 mode or 

press 

     

  key to return to the measurement mode. 

 

P4.3. – Select backlight timing (1-2-3-On) 

1. When press 

   

    key,“

1

” flashes, then press 

   

    key to select blinking 

2

3

On. When the parameter flashes, press          key to confirm. 

Select 

On

 to turn on the backlight, the time unit is minute. 

2. After confirm the parameter, press 

     

  key to enter in P4.4 mode or 

press 

     

  key to return to the measurement mode. 

Summary of Contents for SX811-BS

Page 1: ...SX811 BS Portable pH Meter for Meat Testing Instruction Manual APERA INSTRUMENTS LLC www aperainst com...

Page 2: ...6 4 1 Information regarding pH Calibration 6 4 1 1 Standard Buffer Solution 6 4 1 2 3 Point Calibration 7 4 1 3 How often to calibrate 7 4 1 4 Calibration reminder 7 4 1 5 Check calibration time 8 4 1...

Page 3: ...17 7 7 Data processing 17 8 What s in the box 17 9 Warranty 18 10 Appendix 18 10 1 Appendix I Parameter setup and factory default setup 18 10 2 Appendix II Abbreviation glossary 19 10 3 Appendix III S...

Page 4: ...nced digital processing technology improvs meter responding time and accuracy There are reading stability display mode and automatic lock up display mode n IP57 waterproof operable under extreme condi...

Page 5: ...do not scratch the stainless steel casing and blades of the electrode with sharp object or sand paper 2 Technical Specifications 2 1 Meter s Technical Specs Technical Specs pH Range 2 00 to 19 99 pH...

Page 6: ...ng Life Reference System Electrolyte Polymer Shape of glass membrane Spear Range 0 to 14 pH Temperature Range 0 to 80 Connector BNC and RCA 3 Instrument Description 3 1 LCD Display 1 Measurement mode...

Page 7: ...ture compensation 14 Calibration guide icon 15 Stability icon of readings 16 USB icon when this icon appears the meter connects the computer 17 Low battery icon when this icon appears please renew the...

Page 8: ...ess lIn pH measurement mode press and hold this key to change the resolution repeatedly 0 01 0 1pH Short long press lIn the mode of manual temperature compensation MTC when press and hold this key the...

Page 9: ...inutes the 2nd measuring value is stored See Diagram 8 the meter stores automatically eight measuring values When press key icon stops flashing and the meter stops automatic storage In automatic stora...

Page 10: ...H 4 00 1 68 pH and 9 18 pH 7 00pH 4 00 1 68 pH and 10 01 pH 6 86pH 4 01 1 68pH 9 18 pH 0 to 14 00 pH 4 1 3 How often to calibrate The frequency that you need to calibrate your meter depends on the tes...

Page 11: ...icon displays on LCD Press key to calibrate the meter End icon appears after calibration is done The 1st point calibration is finished In the meanwhile CAL2 will flash at the upper right corner and 4...

Page 12: ...lue flashes Press key or key to adjust the measuring value to 1 60 then press key to calibrate the meter After calibration is done LCD at the top right shows blinking CAL2 icon indicating the meter en...

Page 13: ...m parameter setup P1 6 please see Item 8 3 PU 2 icon displays at the right top of LCD please refer to Diagram 14 Note In parameter setup either PU 1 or PU 2 measurement mode can be selected but both c...

Page 14: ...is to that of the calibration solution the more accurate the readings will be 6 Factory default setting For factory default setting please refer to parameter P1 6 Item 8 3 Per parameter P1 6 all calib...

Page 15: ...g solution for 6 hours c For heavily polluted platinum surface on which there is oxidation film polish the platinum surface with toothpaste then wash it in pure water then submerge it in the soaking s...

Page 16: ...flashes press to confirm 3 After confirm parameter press key to enter in P1 3 mode or press key to return to the measurement mode P1 3 Check the time and date of the last calibration 1 The time and d...

Page 17: ...ult setting 2 When do not select Yes press key to return to the measurement mode 6 4 Submenu of basic parameter setup press key or key to switch P4 1 Adjust timing storage time 1 In mode P4 0 press ke...

Page 18: ...2 After confirm the parameter press key to enter in P4 7 mode or press key to return to the measurement mode P4 7 Adjust date 1 Press key Month flashes then press key Date flashes then press key Year...

Page 19: ...uter pH mV conductivity and dissolved oxygen are classified in the file Export press this key to export the stored value to Microsoft Excel file Exit press this key PC Link program exits from the comp...

Page 20: ...downloaded to the computer pH and mV are sorted in the program 7 6 Storage during operation During operation press key to store or set timing storage The measuring information is downloaded to the com...

Page 21: ...the products purchased by customers not the service life of the tester or probe 10 Appendix 10 1 Appendix I Parameter setup and factory default setup Modes Prompts Parameter setting items Abbreviation...

Page 22: ...off Auto power off P4 5 Clear readings Clear readings P4 6 P4 7 P4 8 Others China China United States of America United States of America Nist Nist Off Off On On No No Yes Yes 10 3 Appendix III Self d...

Page 23: ...20...

Page 24: ...21 APERA INSTRUMENTS LLC Address 977 Old Henderson Rd Columbus Ohio 43220 Tel 1 614 285 3080 Email info aperainst com Website www aperainst com...

Reviews: