- 12 -
2) Clean and activate ORP electrode
After the electrode has been used over a long period of time, the platinum surface will get polluted which
causes inaccurate measurement and slow response. Please refer to the following methods to clean and
activate ORP electrode:
(a) For inorganic pollutant, submerge the electrode in 0.1mol/L dilute hydrochloric acid for 30 minutes,
then wash it in pure water, then submerge it in the soaking solution for 6 hours.
(b) For organic or lipid pollutant, clean the platinum surface with detergent, then wash it in pure water,
then submerge it in the soaking solution for 6 hours.
(c) For heavily polluted platinum surface on which there is oxidation film, polish the platinum surface with
toothpaste, then wash it in pure water, then submerge it in the soaking solution for 6 hours.
6 Parameter Setup
6.1 Main menu
In the measurement mode, press and hold key to enter in mode P1.0, then press
or
to
switch among main menu: P1.0→P2.0→P3.0→P4.0. Please refer to Diagram – 20.
P1.0: pH parameter setup menu,
P4.0: Basic parameter setup menu.
6.2 Submenu
1) In P1.0 mode, press key to enter in submenu P1.1 of pH parameter setup, then press
and key to switch among submenu: P1.1→P1.2→…→P1.6, refer to Diagram – 20.
2) In P4.0 mode, press
key to enter in submenu P4.1 of basic parameter setup, then press
and key to switch among submenu: P4.1→P4.2→…→P4.8, refer to Diagram – 20