2.5

 

Spear pH Electrode is suitable for measure wet solid or semi-solid medium When piercing sample 
medium, do not use excessive force. If the sample medium is too hard (such as meat), please cut a 
small hole before piercing the electrode into it. 

2.6

 

After 1-year of use, we recommend replacing the electrode for the best accuracy.

 

 
 

2. How to Test Soil pH Directly 

a)

 

If soil is dry, add small amount of distilled or RO water to moisten the soil (

do NOT add tap 

water since it will affect the pH value significantly

). Ideally wait 24 hours before you take a 

measurement. 

b)

 

Insert the pH connector (blue one) to the pH socket, and twist it clockwise to lock in; Insert the 
temperature connector to the temperature socket. 

c)

 

Rinse the probe with distilled water/RO water before inserting it to the soil. 

d)

 

If measuring outside of solution, rockwool or small pots, remove the top 5cm/2inch soil from the 
surface of the sample area. 

Insert the probe approximately 4 to 6 inches (10 to 15cm) into the 

soil samples from various locations (3 to 4)

, wait for the stable readings before recording (smiley 

face icon stays on the screen).   

e)

 

Users can press the M+ button to save the stable readings on the meter. 

Then take the average 

of the measured data as the representative pH level of the sample area.

 

f)

 

Do 

NOT

 use extra force to insert probe. If you feel like the probe can’t be inserted any deeper, 

stop inserting. The spear tip could break. 

g)

 

Rinse off the soil residue between tests using distilled water for the best results. Use the probe 
cleaning brush in the case if needed. 

h)

 

Performing calibration of the meter at least every two weeks to ensure accuracy. Refer to section 
3.2 for calibration process. 

i)

 

The connector of the electrode should be kept clean and dry. If contaminated, please clean it with 
medical cotton and absolute alcohol and blow dry to prevent the short circuit of the electrode and 
slow reaction of electrode. 

j)

 

After 1-year of use, we recommend replacing the electrode for the best accuracy. 

 
 
3. Limited Warranty 

We warrant this electrode to be free from defects in material and workmanship and agree to repair or 
replace  free  of  charge,  at  option  of APERA  INSTRUMENTS,  LLC,  any  malfunctioned  or  damaged 
product attributable to responsibility of APERA INSTRUMENTS, LLC for a period of SIX MONTHS 
from the delivery.   
 
This limited warranty does not cover any damages due to: 
Transportation,  storage,  improper  use,  failure  to  follow  the  product  instructions  or  to  perform  any 
preventive  maintenance,  modifications,  combination  or  use  with  any  products,  materials,  processes, 
systems or other matter not provided or authorized in writing by us, unauthorized repair, normal wear and 
tear, or external causes such as accidents, abuse, or other actions or events beyond our reasonable control. 

 
 
 

APERA INSTRUMENTS, LLC 

Address:  6656  Busch  Blvd,  Columbus,  OH  43229                  Phone:  1-614-285-3080 
Website: 

www.aperainst.com

                                                         

Email:  [email protected]

 

Reviews: