background image

 

- 8 - 

soaked in clean water at room temperature for 24 hours. If the water temperature is 

low, soaking time should be 48-72 hours. 

(D)  The sponge can not be allowed to get stained or moldy, otherwise it will consume or 

produce oxygen. If stained or moldy, please replace it immediately (some back-up 

sponges are included in the kit).   

4.3 

Sensor Cap 

(A)  The sensor cap is a key part of the optical DO probe. The surface coating of the cap 

cannot be scratched or mechanically worn. Otherwise, the service life of the sensor 

cap will be reduced or the probe will be damaged. Please pay special attention to it 

when using the probe. 

(B)  The surface coating of the sensor cap cannot withstand high temperature, so the 

optical DO probe can not be tested in water above 50

˚C/122˚F. 

(C)  If the surface of the sensor cap is contaminated, please do not use alcohol or organic 

solvents to clean, otherwise it may damage the probe. It can be gently wiped with a 

soft clean cloth. To disinfect the probe, immerse it in 3% hydrogen peroxide for 15 to 

30 minutes and then rinse off with clean water. 

(D)  The sensor cap has a service life of more than 8000 hours. When the probe is not 

being used, the service life will not be reduced, so the actual use time of the sensor 

cap is far more than a year. The major factor affecting the service life of the sensor 

cap is the surface coating being damaged under external force. So the key is to 

protect the sensor cap from external damage. 

(E)  If the sensor cap is damaged or deteriorated, users need to purchase a new one. 

Every new cap has a set of calibration codes which need to be input into the 

instrument. The specific input method will be described in the instruction manual of 

sensor cap. 

(F)  Users should not take off the sensor cap when it is not in use. Nor should one swap 

the caps from different instruments. When being installed, the sensor cap must be 

tightened, and the interior can not be contaminated or wet.   

 

Preparation for Calibration  

 

5.1 

Dissolved Oxygen Units Selection   

Dissolved oxygen units displays in two forms: mg / L and %, and ppm and %. Press       

   

to   

switch between mg / L → %, or ppm → %. Users can choose mg / L or ppm in parameter 
setting P3.1, but only a percentage (DO %) is displayed in calibration.

   

 

 

Summary of Contents for DO850

Page 1: ...1 DO850 Portable Optical Dissolved Oxygen Meter Instruction Manual APERA INSTRUMENTS LLC aperainst com v3 0...

Page 2: ...ng 6 3 8 Automatic Power Off 6 4 Optical Dissolved Oxygen Probe 7 4 1 Probe Structure 7 4 2 Probe Maintenance 7 4 3 Sensor Cap 8 5 Preparation for Calibration 8 5 1 Dissolved Oxygen Units Selection 8...

Page 3: ...n such as scratches to the light shielding layer the sensor cap has up to 8000 hours of service life Easy to calibrate and maintain Probe is equipped with a calibration storage sleeve which makes cali...

Page 4: ...0 response Calibration Points Saturation Point Zero Oxygen Temperature Compensation Automatic 0 to 50 Pressure Compensation Automatic 60 to 120 kPa Salinity Compensation Manual 0 to 45 ppt Temperature...

Page 5: ...surement mode press the key for 2 seconds to enter the calibration mode Short Press To cancel any operation press to return to measurement mode Short Press In the menu mode press the key to change the...

Page 6: ...n Select On to turn on automatic locking When the reading is stable for more than 10 seconds the meter automatically locks the measured value and displays the HOLD icon as shown in Figure 8 When auto...

Page 7: ...slow The electrode calibration sleeve is used to store the probe A Short term storage less than 30 days The probe head is kept in the calibration sleeve Always keep the sponge inside the calibration s...

Page 8: ...xide for 15 to 30 minutes and then rinse off with clean water D The sensor cap has a service life of more than 8000 hours When the probe is not being used the service life will not be reduced so the a...

Page 9: ...ity of freshwater is 0 to 0 5ppt salinity of seawater is 35ppt 6 Calibration 6 1 Saturated Oxygen Calibration A This procedure requires the use of a calibration sleeve to allow the probe to be calibra...

Page 10: ...he sensor cap can adversely affect the stability of the measurement Please pay special attention to this situation See Section 4 2 Probe Maintenance for details C The instrument has factory default se...

Page 11: ...ppm P3 2 Resolution Selection 0 01 0 1 mg L ppm P3 3 Salinity Compensation 0 to 45 ppt P3 4 Air Pressure Calibration 60 to 120 kPa P3 5 Back to Factory Default No Yes P4 0 Basic parameter P4 1 Temp Un...

Page 12: ...press or to switch P4 1 temperature unit 1 In P4 0 mode Press to enter P4 1 mode 2 Press flashes Press to select Press to confirm 2 Press to enter P4 2 mode or to return to measurement mode P4 2 Auto...

Page 13: ...the date of purchase If the above products or parts within the warranty period is not functional due to raw material defects or poor manufacturing they are free for repair or replace 10 2 Damage and...

Page 14: ...e measured value is not stable prolong the stabilization time until is fully stabilized then press to finish 3 Check the sensor cap If it is contaminated it can be cleaned if dried out it can be hydra...

Page 15: ...7 62 7 22 19 0 9 28 8 79 8 33 7 90 7 48 7 09 20 0 9 09 8 62 8 17 7 75 7 35 6 96 21 0 8 92 8 46 8 02 7 61 7 21 6 84 22 0 8 74 8 30 7 87 7 47 7 09 6 72 23 0 8 58 8 14 7 73 7 34 6 96 6 61 24 0 8 42 7 99...

Page 16: ...82 07 820 73 95 28 43 722 0 96 26 962 59 80 23 94 608 0 81 06 810 60 94 28 13 714 4 95 25 952 46 79 23 64 600 4 80 05 800 47 93 27 83 706 8 94 23 942 32 78 23 34 592 8 79 03 790 34 92 27 53 699 2 93 2...

Reviews: