background image

 

- 11 - 

influenced by temperature, and the probe senses temperature slower than dissolved 

oxygen. 

Parameter Settings 

 
8.1 Press the        key in the measurement mode to enter the parameter setting mode P3.0, 

press the          key to switch the menu P3.0 

→P4.0; In P3.0 mode, press          to enter 

P3.1, press        to switch submenu P3.1

→P3.5; In P4.0 mode, press          to enter P4.1, 

press        to switch submenu P4.1

→P4.4. See Table 2 for details. 

 

 

Table-2 Parameter Setting List 

Menu 

Submenu 

Parameter 

Code 

Content 

P3.0 

DO 

parameter 

P3.1 

DO Units Selection 

mg/L—ppm 

P3.2 

Resolution Selection 

0.01/0.1 mg/L(ppm) 

P3.3 

Salinity Compensation 

(0 to 45) ppt 

P3.4 

Air Pressure Calibration 

 

(60 to 120) kPa 

P3.5 

Back to Factory Default 

 

No

Yes 

P4.0 

Basic 

parameter 

P4.1 

Temp. Unit Selection 

 

P4.2 

Auto Lock 

Off—On 

P4.3 

Auto Backlight 

 

On—Off 

P4.4 

Auto Power Off 

 

On—Off 

 

8.2 DO Parameter setting (press          or          to switch) 

 

  

P3.1—Dissolved Oxygen Unit

 (mg/L—ppm) 

1. In P3.0 mode, press          to enter P3.1 mode.

 

2. Press        , 

mg / L

 flashes, press          to select

   

     

mg/L → ppm, press          to confirm. 

3. Press          to enter P3.2 mode, or press          to return to

 

measurement mode.

 

 

P3.2—Resolution

 (0.01-0.1mg/L) 

1.Press      , , 

0.01

 flashes, press        to select resolution   

(0.01-0.1mg/L), press          to confirm.   

2. Press        to enter P3.3 mode, or press          to return to

 

    measurement mode.

 

Summary of Contents for DO850

Page 1: ...1 DO850 Portable Optical Dissolved Oxygen Meter Instruction Manual APERA INSTRUMENTS LLC aperainst com v3 0...

Page 2: ...ng 6 3 8 Automatic Power Off 6 4 Optical Dissolved Oxygen Probe 7 4 1 Probe Structure 7 4 2 Probe Maintenance 7 4 3 Sensor Cap 8 5 Preparation for Calibration 8 5 1 Dissolved Oxygen Units Selection 8...

Page 3: ...n such as scratches to the light shielding layer the sensor cap has up to 8000 hours of service life Easy to calibrate and maintain Probe is equipped with a calibration storage sleeve which makes cali...

Page 4: ...0 response Calibration Points Saturation Point Zero Oxygen Temperature Compensation Automatic 0 to 50 Pressure Compensation Automatic 60 to 120 kPa Salinity Compensation Manual 0 to 45 ppt Temperature...

Page 5: ...surement mode press the key for 2 seconds to enter the calibration mode Short Press To cancel any operation press to return to measurement mode Short Press In the menu mode press the key to change the...

Page 6: ...n Select On to turn on automatic locking When the reading is stable for more than 10 seconds the meter automatically locks the measured value and displays the HOLD icon as shown in Figure 8 When auto...

Page 7: ...slow The electrode calibration sleeve is used to store the probe A Short term storage less than 30 days The probe head is kept in the calibration sleeve Always keep the sponge inside the calibration s...

Page 8: ...xide for 15 to 30 minutes and then rinse off with clean water D The sensor cap has a service life of more than 8000 hours When the probe is not being used the service life will not be reduced so the a...

Page 9: ...ity of freshwater is 0 to 0 5ppt salinity of seawater is 35ppt 6 Calibration 6 1 Saturated Oxygen Calibration A This procedure requires the use of a calibration sleeve to allow the probe to be calibra...

Page 10: ...he sensor cap can adversely affect the stability of the measurement Please pay special attention to this situation See Section 4 2 Probe Maintenance for details C The instrument has factory default se...

Page 11: ...ppm P3 2 Resolution Selection 0 01 0 1 mg L ppm P3 3 Salinity Compensation 0 to 45 ppt P3 4 Air Pressure Calibration 60 to 120 kPa P3 5 Back to Factory Default No Yes P4 0 Basic parameter P4 1 Temp Un...

Page 12: ...press or to switch P4 1 temperature unit 1 In P4 0 mode Press to enter P4 1 mode 2 Press flashes Press to select Press to confirm 2 Press to enter P4 2 mode or to return to measurement mode P4 2 Auto...

Page 13: ...the date of purchase If the above products or parts within the warranty period is not functional due to raw material defects or poor manufacturing they are free for repair or replace 10 2 Damage and...

Page 14: ...e measured value is not stable prolong the stabilization time until is fully stabilized then press to finish 3 Check the sensor cap If it is contaminated it can be cleaned if dried out it can be hydra...

Page 15: ...7 62 7 22 19 0 9 28 8 79 8 33 7 90 7 48 7 09 20 0 9 09 8 62 8 17 7 75 7 35 6 96 21 0 8 92 8 46 8 02 7 61 7 21 6 84 22 0 8 74 8 30 7 87 7 47 7 09 6 72 23 0 8 58 8 14 7 73 7 34 6 96 6 61 24 0 8 42 7 99...

Page 16: ...82 07 820 73 95 28 43 722 0 96 26 962 59 80 23 94 608 0 81 06 810 60 94 28 13 714 4 95 25 952 46 79 23 64 600 4 80 05 800 47 93 27 83 706 8 94 23 942 32 78 23 34 592 8 79 03 790 34 92 27 53 699 2 93 2...

Reviews: