7.  pH Measurement 

  How to take pH measurements   

Short press 

  to turn on the tester. Rinse the probe in distilled water, 

shake off excess water. Insert the probe in your sample solution, make a 

quick  stir  and  hold  still.  Record  the  reading  when  it  is  stabilized  (
appears and stays on screen). 
 

  Pure Water pH Measurement 

When  testing  pure  water  like  tap  water,  drinking  water,  RO  water  and 

distilled  water,  it  will  take  longer  for  the  readings  to  get  fully  stabilized 

(typically 1-5 minutes). Please be patient. If still not working, add Apera 3M KCL (AI1107) to your 

pure water at the ratio of 1:1000 (e.g. 1 ml KCL to 1000 ml water) to accelerate stabilization while 

minimizing  pH  change.  If  the  accuracy  is  not  meeting  your  requirement,  please  contact  us  at 

[email protected] to find the specialized meter designed for pure water pH test. 

 

8.  Conductivity Calibration 

  How to calibrate 

1) 

Power on the tester. Press 

  key to switch to conductivity measurement mode (

Cond

). Rinse 

the probe in distilled water and dry it. 

2) 

Pour  certain  amount  of  1413μS/cm  and  12.88mS/cm  conductivity  calibration  solution  into 
corresponding calibration bottles (to about half volume of the bottle). 

3) 

Long press 

  key to enter calibration mode, (screen turns green). 

4) 

Dip the probe into 1413 μS/cm conductivity calibration solution, stir gently and hold still. When 

the  reading  is  fully  stabilized  ( 

 stays  on  screen),  short  press 

     to  start  one-point 

calibration, the tester returns to measurement mode and 

  will appear at the bottom left of 

the screen, indicating a successful 1

st

 point conductivity calibration. 

5) 

If needed (your target conductivity level is greater than 2 mS or 2000 μS), insert the probe into 
the 12.88 mS conductivity calibration solution. Follow the steps in 8.1.3 to 8.1.4 to complete the 

2

nd

  point  of  calibration  using  the  12.88  mS  standard  solution. 

   will  appear  next  to 

 , 

indicating a successful 2-point conductivity calibration. 

 

  Conductivity Calibration Notes 

The tester can calibrate 84μS, 1413 μS/cm and 12.88 mS/cm conductivity calibration solution. User 
can conduct 1 to 3 points calibration. Refer to the table below. Usually calibrating the tester with 
1413 μS/cm conductivity buffer solution alone shall meet the testing requirement. 

 
 
 

Summary of Contents for AI3719

Page 1: ...1 SKU AI3719 ZenTest PC60 Z Smart 7 in 1 Tester pH Cond TDS Salinity Resistivity ORP Temp Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Page 2: ...be Replacement 14 17 Troubleshooting Guide 14 18 Warranty 16 ATTENTION Water droplets are added during production to maintain the moisture of the probe This is normal practice and should not be attrib...

Page 3: ...er to the functions available on each platform in the following table This manual shows you how to operate the tester without connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mob...

Page 4: ...an be added to saved data Data Output N A Share data via Email 2 What s in the Kit 3 Battery Installation Please install batteries according to the following steps Please note direction of batteries A...

Page 5: ...tester long press to enter parameter setting 2 In calibration mode or parameter setting short press to return to measurement mode 3 In measurement mode long press to turn off the tester short press to...

Page 6: ...oint calibrations 4 00 10 01 1 68 or 12 45 immediately after the 1st point calibration is finished Do NOT turn off the meter before you calibrate 2nd or 3rd point Otherwise you will need to restart th...

Page 7: ...pH electrode is damaged or broken If so please replace with a new one 2 The glass bulb or junction is severely contaminated Please use a soft brush with soap water to clean it thoroughly Then soak it...

Page 8: ...ur certain amount of 1413 S cm and 12 88mS cm conductivity calibration solution into corresponding calibration bottles to about half volume of the bottle 3 Long press key to enter calibration mode scr...

Page 9: ...tion is expired or contaminated 3 Check if the conductivity electrode two black rods is damaged 4 Check if the conductivity electrode is contaminated If so please use a soft brush with warm water to c...

Page 10: ...ductivity is linear related and its conversion factor is 0 40 to 1 00 Adjust the factor in parameter setting P5 based on the requirements in different industries The factory default setting is 0 71 d...

Page 11: ...I1166 or detergent water for 30 minutes Then use a soft brush to remove the contaminants Afterwards soak the probe in 3M KCL soaking solution for at least 1 hour Rinse it off then re calibrate the tes...

Page 12: ...ding is locked 3 Auto Backlight P03 Users can set the automatic backlight time for 1 to 8 minutes For example if 3 minutes is set the backlight will turn off automatically after 3 minutes when the Off...

Page 13: ...alibration is finished or the reminder setting is cancelled in the ZenTest App the Er6 icon will disappear 6 pH Back to Factory Default P08 and Conductivity Back to Factory Default P12 Select Yes to r...

Page 14: ...ry probe will eventually age and fail even if you don t use it that often The typical service life of Apera probes is 12 24 months depending on the frequency of usage and how well you keep it clean an...

Page 15: ...tabilizing add 1ml of 3M KCL solution to 1000ml of test solution or 1 teaspoon to 1 gallon Display similar readings in any solutions or always display 7 0 pH Broken probe If you don t find any visible...

Page 16: ...PERA INSTRUMENTS Europe GmbH for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to accidental damage unauthorized repair normal w...

Reviews: