6.  pH Calibration 

  Short press 

  to turn on the meter; rinse the probe in distilled 

water, Shake off excess water   

  Pour certain amount of pH 7.00 and pH 4.00 buffer solution into the 

corresponding calibration bottles (to about half volume of the bottle); 

  Long press 

  to enter calibration mode; Short press 

  to exit. 

  Insert the probe into pH 7.00 buffer solution, make a quick stir, and 

hold still. When the reading is stabilized (

stays on the LCD screen), short press 

  to 

complete 1

st

 point calibration. After calibration is completed, the tester will return to 

measurement mode. Icon 

  will appear at the bottom left of the LCD screen. indicating a 

successful 1st point pH calibration. 

  To calibrate 2nd point, use 4.00 pH buffer and repeat Step 6.3 to 6.4 (Do NOT turn off the 

tester after you finish pH 7 calibration). 

  will display next to 

, indicating a successful 2-

point pH calibration (low and middle points).   

  If necessary (target pH>8.00), calibrate 3rd point using 10.01 standard pH buffer and repeat 

Step 6.3 to 6.4, 

  will show up next to 

  and 

, indicating a successful 3-point 

calibration (high, low, and middle points). 

  Notes about Calibrationhe   

1)  The 1st point calibration must be 7.00 pH. Perform the 2nd and 3rd point calibrations (4.00, 

10.01, 1.68, or 12.45) immediately after the 1st point calibration is finished. Do NOT turn off 

the meter before you calibrate 2nd or 3rd point. Otherwise, you will need to restart the 

calibration process with 7.00 pH first.   

2)  The pH 4.00 and 7.00 buffer solutions poured into the calibration vials can be used for up to 10 

times as long as they are not contaminated and the bottles are capped when not in use. pH 

10.01 can only be used for up to 5 times as it will lose its accuracy much faster. After that, 

replace the buffer solutions in the calibration vials with new ones to keep the accuracy. 

Keeping the freshness and cleanliness of calibration buffers is essential for accurate pH 

measurement. 

3)  The tester can perform 1 to 3 points of automatic calibration and can recognize 5 types of pH 

standard solutions. For details, please refer to the following table: 

Calibration 

USA Series 

NIST Series 

Indication icon  Recommended 

1-point     

7.00 pH 

6.86 pH 

 

Accuracy 

requirement ≥ 0.1 

pH 

2-pt 

Option 

1

st

 pt: 7.00 pH 

2

nd

 pt: 4.00 pH or 

1.68 pH 

Option 

1

st

 pt: 6.86 pH 

2

nd

 pt: 4.01 pH or 1.68 

pH 

 

Range < 7.00 pH 

Option 

1

st

 pt: 7.00 pH 

2

nd

 pt: 10.01 pH or 

12.45 pH 

Option 

1

st

 pt: 6.86 pH 

2

nd

 pt: 9.18 pH or 

12.45 pH 

 

Range >7.00 pH 

3-pt 

1

st

 pt: 7.00 pH 

2

nd

 pt: 4.00 or 1.68 pH 

3

rd

 pt: 10.01 or 12.45 pH 

1

st

 pt: 6.86 pH 

2

nd

 pt: 4.01 or 1.68 pH 

3

rd

 pt: 9.18 pH or 12.45 pH 

 

 

Range: 0 to 14.00 

pH 

Summary of Contents for AI3719

Page 1: ...1 SKU AI3719 ZenTest PC60 Z Smart 7 in 1 Tester pH Cond TDS Salinity Resistivity ORP Temp Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Page 2: ...be Replacement 14 17 Troubleshooting Guide 14 18 Warranty 16 ATTENTION Water droplets are added during production to maintain the moisture of the probe This is normal practice and should not be attrib...

Page 3: ...er to the functions available on each platform in the following table This manual shows you how to operate the tester without connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mob...

Page 4: ...an be added to saved data Data Output N A Share data via Email 2 What s in the Kit 3 Battery Installation Please install batteries according to the following steps Please note direction of batteries A...

Page 5: ...tester long press to enter parameter setting 2 In calibration mode or parameter setting short press to return to measurement mode 3 In measurement mode long press to turn off the tester short press to...

Page 6: ...oint calibrations 4 00 10 01 1 68 or 12 45 immediately after the 1st point calibration is finished Do NOT turn off the meter before you calibrate 2nd or 3rd point Otherwise you will need to restart th...

Page 7: ...pH electrode is damaged or broken If so please replace with a new one 2 The glass bulb or junction is severely contaminated Please use a soft brush with soap water to clean it thoroughly Then soak it...

Page 8: ...ur certain amount of 1413 S cm and 12 88mS cm conductivity calibration solution into corresponding calibration bottles to about half volume of the bottle 3 Long press key to enter calibration mode scr...

Page 9: ...tion is expired or contaminated 3 Check if the conductivity electrode two black rods is damaged 4 Check if the conductivity electrode is contaminated If so please use a soft brush with warm water to c...

Page 10: ...ductivity is linear related and its conversion factor is 0 40 to 1 00 Adjust the factor in parameter setting P5 based on the requirements in different industries The factory default setting is 0 71 d...

Page 11: ...I1166 or detergent water for 30 minutes Then use a soft brush to remove the contaminants Afterwards soak the probe in 3M KCL soaking solution for at least 1 hour Rinse it off then re calibrate the tes...

Page 12: ...ding is locked 3 Auto Backlight P03 Users can set the automatic backlight time for 1 to 8 minutes For example if 3 minutes is set the backlight will turn off automatically after 3 minutes when the Off...

Page 13: ...alibration is finished or the reminder setting is cancelled in the ZenTest App the Er6 icon will disappear 6 pH Back to Factory Default P08 and Conductivity Back to Factory Default P12 Select Yes to r...

Page 14: ...ry probe will eventually age and fail even if you don t use it that often The typical service life of Apera probes is 12 24 months depending on the frequency of usage and how well you keep it clean an...

Page 15: ...tabilizing add 1ml of 3M KCL solution to 1000ml of test solution or 1 teaspoon to 1 gallon Display similar readings in any solutions or always display 7 0 pH Broken probe If you don t find any visible...

Page 16: ...PERA INSTRUMENTS Europe GmbH for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to accidental damage unauthorized repair normal w...

Reviews: