ATTENTION

 

  Water droplets are added during production to maintain the moisture of the probe. This is 

normal practice and should not be attributed to used product. 

 

Never

 

use the product when it’s freezing cold. Let it warm to room temperature before using.

 

 

 

 

 

 

 

 

 

Summary of Contents for AI3715

Page 1: ...1 ZenTest PH60F Z Smart Flat pH Tester pH Temp ORP Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Page 2: ...Storage 8 11 Parameter Setting 9 12 Technical Specifications 10 13 Icons and Functions 10 14 Probe Replacement 11 15 Troubleshooting Guide 11 16 Warranty 12 ATTENTION Water droplets are added during...

Page 3: ...ut connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LCD display 1 Basic Mode digital display calibration info Swipe to...

Page 4: ...on Please install batteries according to the following steps Please note direction of batteries All POSITIVE SIDES FACING UP Wrong installation of batteries will cause damage to the tester and potenti...

Page 5: ...urn on the meter rinse the probe in distilled water Shake off excess water 1 When turned off short press to turn on the tester long press to enter parameter setting 2 In calibration mode or parameter...

Page 6: ...after the 1st point calibration is finished Do NOT turn off the meter before you calibrate 2nd or 3rd point Otherwise you will need to restart the calibration process with 7 00 pH first 2 The pH 4 00...

Page 7: ...ed Please use a soft brush with soap water to clean it thoroughly Then soak it in 3M KCL overnight before performing calibration again 3 The electrode is aged used for over a year and has a much slowe...

Page 8: ...d water and dry it Dip the probe in sample solution shake for a few seconds and allow it to stand still Get the ORP readings after appears and stays on screen 9 Probe Cleaning 1 The tester is only as...

Page 9: ...isplayed when reading is locked 3 Auto Backlight P03 Users can set the automatic backlight time for 1 to 8 minutes For example if 3 minutes is set the backlight will turn off automatically after 3 min...

Page 10: ...Icons and Functions pH Range 2 00 to 16 00 pH Resolution 0 01 pH Accuracy 0 01 pH 1 digit Calibration Points 1 to 3 points Auto Temperature Compensation 0 50 C 32 122 F ORP mV Range 1000 mV to 1000 mV...

Page 11: ...ng Incorrect standard solutions Reboot tester calibrate pH 7 first then pH 4 For details refer to Section 5 2 a Poor quality standard solutions Replace with fresh and clean standard calibration soluti...

Page 12: ...ean the probe with cleaning solution then soak it in 3M KCL soaking solution overnight Comparison with other testers test strips or drop tests To compare with other testers make sure to perform a 2 po...