Apera Instruments 28APE190030/AI5563 User Manual Download Page 9

 

Screen 

3-color LCD screen, 

Blue: Measurement; Green: Calibration; Red: Alarm 

Reading Lock 

HOLD 

Low-Voltage Warning 

  flashing, reminder of battery replacement needed 

Auto. Power-Off 

In 8 minutes without operation 

Waterproof Rating 

IP67 

Power 

DC3V, AAA alkaline batteries×4 

Battery Life 

Operation up to 2000 hours 

Dimension & Weight 

Tester: 40×40×178mm/133g; Case: 255×210×50mm/700g 

 

12. Probe Replacement 

12.1  Every pH probe gradually loses its sensitivity and will eventually fail. A typical service life of a 

pH probe is 1-2 years depending on many factors such as frequency of use, nature of test 

samples, and how well it is maintained, etc. Apera Instruments recommends replacing the pH 

probe every 1 to 2 years to guarantee the optimal performance. 

 

12.2  To replace a probe: 1) Take off the probe cap; 2) Screw off the 

probe ring; 3) Unplug the probe; 4) Plug in the new replacement 
probe (pay attention to the probe’s position); 5) Screw on the 
probe ring tightly. Soak the probe in 3M KCL for 5-15 minutes. 

Then perform calibration before testing.  

 

12.3  The replacement probes compatible with PH5F:  

  AI1201 PH60-E (Regular pH glass bulb probe for general water solutions) 

  AI3711-E PH60-DE (Double-junction pH glass bulb probe for complex solutions) 

  AI1205 PH60S-E (Spear pH probe for soft-solids pH testing) 

  AI1203 PH60F-E (Flat pH probe for surface pH testing) 

  AI1207 ORP60-E (ORP probe) 

 

13. Notes 

a)  Avoid testing in very high (> 113 °F/ 45°C) or very low (< 41 °F/ 5°C) temperature solutions as 

it will cause greater measurement error and potential damage to the probe. Test your samples 

and perform calibration close to room temperature as much as possible. 

b) 

NEVER 

store the probe in pure water like tap, RO, distilled, or deionized water as they could

 

damage the pH probe. If this happens, immediately soak the pH probe Apera 3M KCL soaking 

solution overnight, then re-calibrate it before using. Pure water is only for rinsing the probe. 

Summary of Contents for 28APE190030/AI5563

Page 1: ...1 Premium Series PH5F pH Tester User Manual APERA INSTRUMENTS Europe GmbH www aperainst de v7 0...

Page 2: ...4 5 pH Measurement 5 6 Probe Cleaning 6 7 Probe Storage 6 8 Parameter Setting 6 9 ORP Measurement 8 10 Technical Specifications 8 11 Other Specifications 8 12 Probe Replacement 9 13 Notes 9 14 Battery...

Page 3: ...parameter setting 3 In measurement mode short press to turn on backlight 4 In calibration mode short press to cancel calibration 1 In measurement mode short press to switch between pH and mV 2 In sett...

Page 4: ...probe in the pH 7 00 buffer solution first and make a quick stir in the solution then hold still 4 1 4 Long press to enter calibration mode screen will turn green Short press if you decide to quit cal...

Page 5: ...points of automatic calibration and can recognize 5 types of pH standard solutions For details please refer to the following table Calibration USA Series NIST Series Indication icon Recommended 1 pt 7...

Page 6: ...after each measurement with pure water in a container or with a wash bottle 6 2 For tough contaminants detach the sensor shield soak the probe in Apera probe cleaning solution AI1166 or detergent wate...

Page 7: ...wo options of standard buffer solutions USA series and NIST series Factory default is USA series for details see clause 4 2 8 3 2 Alarm Function P2 P3 Examples Alarm triggered when measurement 3 20 pH...

Page 8: ...surement ORP stands for Oxidation Reduction Potential ORP is a measure of the cleanliness of the water its ability to break down contaminants An ORP probe is needed to test ORP sold separately SKU AI1...

Page 9: ...he new replacement probe pay attention to the probe s position 5 Screw on the probe ring tightly Soak the probe in 3M KCL for 5 15 minutes Then perform calibration before testing 12 3 The replacement...

Page 10: ...r s cap while making sure to push all the way down The tester s waterproof design may be compromised if the cap is not fitted correctly 15 Troubleshooting Guide Trouble Reason How to fix Cannot calibr...

Page 11: ...robe Replace the probe Air bubbles around the sensor shield Make a quick stir in the solution to remove air bubbles Clogged junction Clean the probe with probe cleaning solution then soak it in 3M KCL...