Trouble 

Reason 

How to fix 

Reading is always 

slowly changing, 

won’t stabilize 

Contaminated probe 

Use a soft brush to clean the probe with Apera probe 

cleaning solution or detergent water. 

Clogged junction 

Use a soft brush to clean the probe with Apera probe 

cleaning solution or detergent water, then soak it in 

Apera 3M KCL soaking solution overnight. 

Aged probe 

Replace the probe. 

Testing pure water pH like 

tap/drinking/RO/distilled water 

Be patient, wait for 1-5 minutes to reach a fully 

stabilized reading. If still not stabilizing, add Apera 3M 

KCL solution to test water at 1:1000 ratio. 

Display similar 

readings in any 

solutions or always 

display 7.0 pH 

Broken probe 

If you don’t find any visible damage of the probe, 

contact us for warranty fulfillment; If there is visible 

damage, replace the probe. 

Instrument defect 

Contact us for warranty fulfillment. 

Reading keeps 

jumping 

Probe is not fully submerged 

in the solution 

Make sure the probe is fully immersed in the solution at 

least 1 inch. 

Air bubbles around the sensor 

shield 

Make a quick stir in the solution to remove air bubbles. 

Probe is not properly 

connected, or the pin 

connector is broken. 

Check the 

probe’s connector, make sure it’s not broken 

and is correctly connected. Align the probe and 

instrument correctly before plugging in. Never force it. 

Ensure that the probe connector is not exposed to air 

too long. 

Calibration is 

successful, but 

reading is not 

accurate 

Aged probe 

Replace the probe. 

Air bubbles around the sensor 

shield 

Make a quick stir in the solution to remove air bubbles. 

Clogged junction 

Clean the probe with probe cleaning solution, then soak 

it in 3M KCL soaking solution overnight. 

Comparison with other 

testers, test strips, or drop 

tests 

To compare accuracy with other testers, make sure to 

perform calibration for all testers in the same standard, 

then test another standard. Whichever gives more 

accurate reading is the more accurate one. Test strips 

or drop tests’ accuracy in not comparable to pH meters‘. 

Poor quality standard 

solutions 

Replace with fresh and clean standard calibration 

solutions made by legitimate scientific instrument 

manufacturers. 

The probe is not suitable for 

your application. 

Contact us to find the most appropriate product for your 

specific application. 

16. 

Warranty

 

We warrant this instrument to be free from defects in material and workmanship and agree to repair 
or replace free of charge, at option of APERA INSTRUMENTS (Europe) GmbH, any malfunctioned 
or damaged product attributable to responsibility of APERA INSTRUMENTS (Europe) GmbH for a 
period of TWO YEARS (SIX MONTHS for the probe) from the delivery.  
This limited warranty does NOT cover any damages due to: 
Accidental damage, transportation, storage, improper use, failure to follow the product instructions 
or to perform any preventive maintenance, unauthorized repair or modifications, normal wear and 
tear, or other external causes or actions beyond our reasonable control. 
 

APERA INSTRUMENTS (Europe) GmbH

 

Wilhelm-Muthmann-Straße 18, 42329 Wuppertal, Germany

 

www.aperainst.de | 

[email protected]

 | Tel. +49 202 51988998 

Summary of Contents for 28APE190030/AI5563

Page 1: ...1 Premium Series PH5F pH Tester User Manual APERA INSTRUMENTS Europe GmbH www aperainst de v7 0...

Page 2: ...4 5 pH Measurement 5 6 Probe Cleaning 6 7 Probe Storage 6 8 Parameter Setting 6 9 ORP Measurement 8 10 Technical Specifications 8 11 Other Specifications 8 12 Probe Replacement 9 13 Notes 9 14 Battery...

Page 3: ...parameter setting 3 In measurement mode short press to turn on backlight 4 In calibration mode short press to cancel calibration 1 In measurement mode short press to switch between pH and mV 2 In sett...

Page 4: ...probe in the pH 7 00 buffer solution first and make a quick stir in the solution then hold still 4 1 4 Long press to enter calibration mode screen will turn green Short press if you decide to quit cal...

Page 5: ...points of automatic calibration and can recognize 5 types of pH standard solutions For details please refer to the following table Calibration USA Series NIST Series Indication icon Recommended 1 pt 7...

Page 6: ...after each measurement with pure water in a container or with a wash bottle 6 2 For tough contaminants detach the sensor shield soak the probe in Apera probe cleaning solution AI1166 or detergent wate...

Page 7: ...wo options of standard buffer solutions USA series and NIST series Factory default is USA series for details see clause 4 2 8 3 2 Alarm Function P2 P3 Examples Alarm triggered when measurement 3 20 pH...

Page 8: ...surement ORP stands for Oxidation Reduction Potential ORP is a measure of the cleanliness of the water its ability to break down contaminants An ORP probe is needed to test ORP sold separately SKU AI1...

Page 9: ...he new replacement probe pay attention to the probe s position 5 Screw on the probe ring tightly Soak the probe in 3M KCL for 5 15 minutes Then perform calibration before testing 12 3 The replacement...

Page 10: ...r s cap while making sure to push all the way down The tester s waterproof design may be compromised if the cap is not fitted correctly 15 Troubleshooting Guide Trouble Reason How to fix Cannot calibr...

Page 11: ...robe Replace the probe Air bubbles around the sensor shield Make a quick stir in the solution to remove air bubbles Clogged junction Clean the probe with probe cleaning solution then soak it in 3M KCL...

Reviews: