![APAR AR234 User Instruction Download Page 4](http://html1.mh-extra.com/html/apar/ar234/ar234_user-instruction_2951921004.webp)
4
protection against incorrect battery polarity;
available accessories:
−
a lithium battery, 3.6 V, type AA (R6), 2450 mAh (e.g. SAFT LS14500);
−
an SD memory card (2 GB);
−
SD/MMC card reader;
−
a stabilized power supply, 5 V/150 mA;
NOTE:
−
before you start working with the recorder, read this instruction, perform the actions
described in chapters
7
and 8
, and then
properly connect the sensor or the input signal and configure the parameters (chapter 9);
−
recording is performed only in a predefined interval (Start and Stop parameters, Table 9, chapter 9).
3. CONTENTS OF THE SET
a recorder
with a 3.6 V AA lithium battery;
a USB (A4 – mini B) cable for connecting the device to a computer, 2 m long;
a CD with the drivers and the software (Windows XP/Vista/7/8/10)
a user instruction;
a warranty card.
4. TECHNICAL DATA
A universal input
(programmable):
measurement range
- Pt100 (RTD, 3- or 2-wire)
-200 ÷ 850
°C
- Ni100 (RTD, 3- or 2-wire)
-50
÷ 170
°C
- thermocouple J
-40
÷ 800
°C
- thermocouple K
-40
÷ 1200
°C
- thermocouple S
-40
÷ 1600
°C
- thermocouple B
300 ÷ 1800
°C
- thermocouple R
-40
÷ 1600
°C
- thermocouple T
-25
÷ 350
°C
- thermocouple E
-25
÷ 680
°C
- thermocouple N
-35
÷ 1300
°C
- current (R
in
= 110 Ω)
0/4
÷ 20
mA
- voltage (R
in
= 110 kΩ )
0
÷ 10
V
- voltage (R
in
> 2 M Ω)
0
÷ 60
mV
- resistance (3- or 2-wire)
0
÷ 700
Ω
Number of measurement inputs
1
Resistance of leads
(RTD, Ω)
R
d
< 25 Ω (for each line)
Resistance input current
(RTD, Ω)
~480 μA (impulse)
Processing errors
(at ambient temperature of 25 °C):
- basic
- for RTD, mA, V, mV, Ω
0.1% of the measurement range ±1 digit
- for thermocouples
0.2% of the measurement range ±1 digit
!