- 8 -

11. LCD DISPLAY OPERATION MODES

The display can be set within one of two operating modes: economic (12:

LCDmod 

= economic) or constant 

(

LCDmod 

= constant). In economic mode the display turns on for a few seconds after performing every 

measurement while recording or can be enabled manually by pressing the "

COPY

" button. This operating 

mode ensures the lowest energy consumption, which allows longer recorder operation without battery 
replacement.
In constant mode, the display is constantly on and updated in the following conditions:
- measurement and write operation occurred during recording (according to 3:

MPer 

parameter, 

  see section 7, Table 1)
- measurement occurred according to the display update period (13:

LCDupd

)

- measurement triggered by pressing the „

COPY

" button

The display update period (13:

LCDupd 

parameter) allows perform measurements also when recording is 

turned off. Set this parameter value over 1 min. to minimize battery energy consumption.

12. MEASSAGES AND ERRORS — LED INDICATORS

Two LED indicators localized on one side of the recorder indicate the occurrence of the following events:

READ/WRITE

 : - read/write to internal storage or SD/MMC

STATUS

:

- transmission via virtual COM port (used by ARSOFT-WZ1 application, see section 7, pt 1)
- copy configuration file from SD/MMC to recorder and move files with recorded data to a 
  card by pressing the „

COPY”

 button (see section 8)

- storage used to save recorded data in an archive file is full - short blink while trying to 
  save the data and the 

F

W

L

L

 message on the display

- battery voltage too low (below 3.15 V) - short blink during measurement output recording 
  and the 

L

b

A

t

 message on the display (additionally a record is created in the archive file, 

  section 9)

13. IMPORTANT OPERATIONAL NOTES

Read this user manual carefully to ensure easy usage of the product. For maximum durability of batteries 
make sure that:
- measurement and recording period parameters are set to the highest possible value (above 10 min. is  
  recommended, parameter 3:

MPer

, Table 1) 

- SD/MMC card is used only when necessary. Otherwise use internal memory
- the display is working in economic mode (parameter 12:

LCDmod

). If constant mode is required, update   

  period should be set to the highest possible value (above 1 min. recommended, parameter 13:

LCDupd

)

- alerts are turned off when not in use (high and low alerts should be set at bounds of the measurement  
  range, section 7, table 1, parameters 14:

AHiHum

,15:

ALoHum

, 16:

AHiTem

, 17:

ALoTem

)

- unnecessary files are removed from the internal memory and SD card before recording starts 
In the following conditions:
- measurement and recording periods are much shorter than 10 min
- SD/MMC card is in use
- the display works in the constant mode (with measured results frequently updated)
consider using optional AC adapter, which may extend durability of batteries up to about 8 years.
Another important thing to note is that you must not disconnect the device from the computer while 
communicating over USB interface - this is signaled by illumination of 

READ/WRITE

 and 

STATUS 

LED 

indicators. USB communication can be defined as the act of working with mass memory device (internal 
SD/MMC card) or running ARSOFT-WZ1 application.
When response from the recorder seems to be suspended remove battery for a while and check its voltage. 
If that voltage is less then 3,3V then replace the battery for a new one of a proper type.

Summary of Contents for AR232

Page 1: ...9 Warsaw ul Klobucka 8 pav 119 Poland Tel 48 22 853 48 56 853 49 30 607 98 95 Fax 48 22 607 99 50 E mail info apar pl Web site www apar pl TEMPERATURE AND HUMIDITY RECORDER AR236 TEMPERATURE RECORDER AR232 Rev 1 1 0 2010 02 16 ...

Page 2: ...L PARTS IDENTIFICATION 4 5 CONNECTING TO THE COMPUTER AND DRIVER INSTALLATION 4 6 SOFTWARE INSTALLATION 4 7 PARAMETERS ADJUSTMENT 5 8 OPERATION AND FEATURES OF SD MMC CARD 6 9 VIEWING RECORDED MEASUREMENTS AND EVENTS 7 10 ALERT CONFIGURATION 7 11 LCD DISPLAY OPERATION MODES 8 12 MESSAGES AND ERRORS LED INDICATORS 8 13 IMPORTANT OPERATIONAL NOTES 8 14 REPLACING THE BATTERY 9 The producer reserves t...

Page 3: ...n SD MMC card n checksum to detect unauthorized changes in the archive n uniquely identifying multiple recorders of the same type by assigning unique identifier ID to each of them n low and high value alerts in band and out of band alert LED indicators n programmable measurement period start and end point of recording and other configuration parameters measured signal zero adjustment COPY button l...

Page 4: ...lithium battery 3 6 V typeAA R6 2450 mAh SAFTLS14500 Working time on new battery 1 up to 5 years in ambient temperature 20 30 C Nominal operation environment 20 70 C 100 RH without condensation Protection level IP20 Casing on wall material ABS UL94 V0 white Casing dimensions 80 x 80 x 25 mm Working position any Weight 97g with battery 1 working time depends on measurement period whether SD MMC car...

Page 5: ... and SD MMC storage available after inserting a card into the SD MMC CARD slot In internal memory is available a configuration text file AR236 cfg or AR232 cfg section 7 NOTE Do not disconnect the device from the computer while installing drivers or when detecting the discs after connecting the recorder to the USB interface READ WRITE indicator illuminated 6 SOFTWARE INSTALLATION The CD shipped wi...

Page 6: ...values from an existing file save the created configuration in a file AR236 cfg or AR232 cfg and store it on an SD MMC card using the card reader insert the card in the SD MMC CARD slot and press COPY while copying the STATUS indicator will be illuminated when the STATUS LED turns off remove the card from the slot NOTE for the off line configuration to be successfully completed parameters for data...

Page 7: ...n which requires removing batteries for a while Notes 1 parameter available only in ARSOFT WZ1 2 when connected to the USB interface of computer the device does not record data in a file 3 the measurement and recording period start with disconnection from USB 4 see section 11 for details about operating modes of the display 5 parameters enabling offset for measurement eg in case discrepancies betw...

Page 8: ... 6 LVBAT x xx where x xx is battery voltage V Use ARSOFT WZ3 application to produce and print graphical or textual representation of the output Moreover the application also enables detection of unauthorized changes of the archive Alternatively the csv files may be edited with spreadsheet software such as Microsoft Excel or OpenOffice Calc as well as word processors e g Windows WordPad or Notepad ...

Page 9: ...rement output recording and the L b A t message on the display additionally a record is created in the archive file section 9 13 IMPORTANT OPERATIONAL NOTES Read this user manual carefully to ensure easy usage of the product For maximum durability of batteries make sure that measurement and recording period parameters are set to the highest possible value above 10 min is recommended parameter 3 MP...

Page 10: ...ace battery with a new one of correct type section 3 make sure the battery is placed correctly in the slot put the circuit board back in place and close the back cover NOTE During battery replacement the real time clock RTC is cleared and needs to be reset with ARSOFT WZ1 application The other parameters remain unchanged Fig 2 Removing circuit board catches Fig 1 Removing back cover ...