![APAR AR232 User Manual Download Page 7](http://html1.mh-extra.com/html/apar/ar232/ar232_user-manual_2951936007.webp)
- 6 -
8. OPERATION AND FEATURES OF SD/MMC CARD
SD/MMC card to store data may be particularly useful for stationary (on-wall) installations of the recorder
and when the amount of measurement data to be stored exceeds the capacity of the internal storage.
Properly installed card in the "
SD/MMC CARD
" slot has the following features:
- store files with values written to the files during recording
- off-line parameter configuration (using a configuration file
AR236.cfg
or
AR232.cfg
with "
COPY
" button,
see section 7)
- move archive
„csv”
files from internal storage onto the card (the action may be performed by pressing
"
COPY
" button and authorizing or when authorization request is disabled; during the operation
"
STATUS
" indicator is illuminated; timing for 4 MB of data - up to 2 min.)
NOTE
:
- the action of moving
„csv”
files from internal storage to SD/MMC card is performed
CONCURRENTLY
with off-line configuration. Thus, to avoid undesired recorder reconfiguration, make sure that on the card
exists appropriate "
.cfg
" file (or the file does not exist if
Prot
=
disabled
)
- use only branded SD/MMC cards (e.g. SanDisk or Kingston) formatted using FAT 16 file system for
maximum battery durability
-
cards with capacity over 2 GB may cause device suspension, which requires removing batteries
for a while
Notes:
(1)
- parameter available only in ARSOFT-WZ1
(2)
-
when connected to the USB interface of computer, the device does not record data in a file
(3)
-
the measurement and recording period start with disconnection from USB
(4)
-
see section 11 for details about operating modes of the display
(5)
- parameters enabling offset for measurement (eg. in case discrepancies between real and
measured values are detected)
(6)
- when
Prot
=
enabled
only SD/MMC cards containing a configuration file with password
(
Pass
) matching the password in recorder will be supported. This prevents unauthorized
removal of archive „
csv
” files stored in internal storage and changing configuration
parameters of the recorder. For
Prot
=
disabled
all SD/MMC cards are supported.
(7)
- parameter available only in ARSOFT-WZ1 application. It is used to identify files with data
created by many different recorders of the same type
(8)
-
see section 10 for details about alert configuration
!
12
LCDmod
display operating mode
(
4
)
economic
or
constant
econ.
13
LCDupd
display update time
1
0
s
÷
24h
, interval 10s
2min
in constant mode
(
4
)
14
AHiHum
high alert for humidity
(
8
)
0
÷
100.0
%RH (AR236 only)
100%RH
15
ALoHum
low alert for humidity
(
8
)
0
÷
100.0
%RH (AR236 only)
0 %RH
16
AHiTem
high alert for temperature
(
8
)
-40.0
÷
100.0
°C
100 °C
17
ALoTem
low alert for temperature
(
8
)
-40.0
÷
100.0
°C
-40 °C
# Name Description Value and value range factory custom
Settings