background image

 Automatikprodukter

3

Calibration (CO2)

IMPORTANT

This CDH  has been factory calibrated and should

      only need calibration once every 12 months. For the most

accurate field calibration possible we recommend

calibrating this sensor with the AP calibration Kit (part

number CD CAT) available from AP.

In the CDH Monitor, any sensor drift usually occurs at it zero

refernce point. The manual calibration process allows the user

to perform a one point calibration based on ambient levels or

by flowing a gas of a known concentration through the sensor.

This process will adjust the zero offset of the sensor and will

provide an accurate calibration. If for reason a two point

calibration is required the calibration kit or CO

2

View

TM

software should be used. To manually calibrate the Model

CDH CO

2

 sensor see instructions below.

1. The calibration procedure will last approximately

15 minutes. Before performing the calibration procedure,

remove the battery cover (see “Battery Installation”) to

provide access to the CO

2

 calibration activation

switch.

Connect the supplied AC adapter to the back of

the sensor.

Ensure the sensor has new batteries or is powered by the

plug in transformer.

2. Power up the sensor and wait for the Warm-up to end.
3. Next verify the Elevation correction has been set.

Refer to the steps in Elevation Correction for procedure.

4. If you are calibrating to ambient conditions make sure the

sensor is displaying a stable reading. It is also very

importent to avoid breathing in the area of the monitor.

If you are flowing gas to the calibration port of the sensor,

allow the gas to flow for at least 10 minutes before initia

ting calibration.

5. Press the 

Mode 

button twice. The Calibration mode will

begin blinking.

6. Press 

Ente

r.

7. Use the Up/Down Rocker button to adjust the lower

reading on the display to the current ambient conditions.

Pressing the button once will change the readings in

increments of 10 ppm.

To increase the speed, press and

hold the button.

NOTE

For best accuracy, a reference or known concentration of CO

2

should be used when adjusting the reading.

8. Next, on the backside of the unit locate the push button

switch (under the battery cover, in the small round hole to

the right of the connector jack), use a small pointed object

to depress and hold the switch for 5 seconds.

The 

Calibration 

light will begin to blink.

9. Press Enter.
10. 

Calibration In Progress 

will begin to blink.

Temperature Adjustment

Use this procedure to adjust the temperature output when, for

example, you wish to have the temperature output match a

reference sensor.

The accuracy of a field adjustment is dependent upon the

stability of the environment is which the procedure is perfor-

med, and upon the accuracy of the reference sensor.
1. Before performing the temperature adjustment, connect

the supplied AC adapter to the back of the sensor. If

you do not have the AC adapter, new batteries should

be used.

2. Power up the sensor and, using the kickstand, place it

on a flat surface in a stable environment relatively free

of drafts or temperature changes. Wait 30 minutes for

the unit to fully equilibrate with the environment. Do not

hold the unit in your hand during this period.

3. Press the 

Mode 

button a number of times. until the

blinking word “TEMPERATURE” appears.

4. Press 

Ente

r. Both the word “TEM-PERATURE” and the

numeric temperature display will begin blinking in unison.

5. Use the Up/Down Rocker button to adjust the temperature

reading to match the reference.

6. Press 

Ente

r. The temperature offset is immediately

adjusted, the blinking stops, and the unit is now in normal

operating mode.

accurately display current room conditions.

Kickstand - 

The sensor is equipped with a kickstand which

can be used for desktop monitoring.

Using an External Datalogger 

- Voltage outputs for both CO

2

and temperature are available via an RJ-45 jack on the rear of

the unit.

Ventilation Rate Indicator (cft)

Overwiew - To display the ventilation rate scroll through the

up/down button until the lower display provides a reading

showing ftm, which stands for cfm (cubic feet per minute). This

value represents how much outside air is being introduced on

a cfm per person basis in the space. Current codes and

standards generally require 15 to 20 cfm/person to be

delivered to most spaces to ensure acceptable indoor air

quality. Lower cfm/person values will indicate lower levels of

ventilation and potentially poor air quality. Higher levels will

indicate excessive ventilation and potential excessive energy

usage. Accurate interpretation of the ventilation rate indicator

requires a measurement be take 2 to 3 hours after occupancy

has stabilized in a space or at a peak in daily CO2

concentrations. In other conditions the indicator may tend to

over estimate ventilations rates. The calculation of ventilation

rates also assumes an office type of activity level (sitting some

walking and other low level activity). If activity levels are

higher the sensor will also underestimate the ventilation rate.
The CDH  will calculate the outside air ventilation rate to a

space based on the inside/outside CO2 differential readings.

The sensor is factory set to assume an outside level of 400

ppm, which should be close to the outside concentration in

most areas. The outside level of CO2 can also be changed by

measuring outside levels or by manually adjusting the

monitor.

Adjustment 

-  To measure readings, take the unit outside and

wait for levels to stabilize on the CO

2

 display (should take

approximately 5 minutes). Once levels have stabilized, hold

the ”enter” button on the sensor for 5 seconds.

At this point the unit will program itself based on the CO

2

value that was input in Step 6.
The calibration process will take approximately 5 minutes.
When Calibration is complete, the display will return to the

steady 

Calibration 

mode.

Press 

Enter 

to return to the normal operation mode.

Reviews: