background image

Automatikprodukter

2

Operation

Power-Up

1. Press the Power Button.
2. A 2 second delay will occur before the display

becomes visible.

3. 10 seconds will elapse before current CO

2

 readings

are displayed.

4. 

“Warm-up” 

will display for approximately one minute.

During this one minute warm-up, adjustments can not

be made to the sensor.

Elevation Correction 

- The sensor is shipped with the

elevation setting set at “zero” or sea level. The sensor, like any

other gas measuring device is directly affected by altitude

changes. If you are at an altitude greater than 500 feet, an

adjustment should be made to assure maximum sensor

accuracy. The altitude correction is very small for changes of

500-1000 ft. (e.g. 10-15 ppm/500 ft. increment). Once the

elevation correction is set it will be stored in the monitor

memory and will be retained if the unit is turned off or batteries

removed.
To change the default setting using Elevation Correction

follow the steps below.
1.  While in Normal Mode press the “Mode” button once.

The Elevation LCD will begin blinking.

2. Press Enter.
3. Press mode to toggle the elevation reading between

feet (ft) and meters (m)

4. Use the Up/Down button to adjust the alti-tude. The

altitude can be adjusted in increments of 500 (feet) or

100 (meters). Once the correct altitude is set, press
Enter to lock the setting and return to normal mode.

Temperature or Cfm/person - 

The lower digital display will

cycle through the display of the following units when up/down

buttons are pressed: Temp°F, Temp°C, cfm/person(fm), turn

lower display off.

Stand-alone Monitoring - 

Once the batteries have been

installed and the Elevation correction has been made (as

described in the steps above), the sensor will begin to

Start-Up

Battery Operation 

- For portable use, the monitor operates

on 4 AA batteries.

Expected battery life is as follows:

Normal 

(Non-Alkaline)....

Up to 50 hours

Rechargeable

..................

Up to 60 Hours

Industrial Alkaline

 ..........

Up to 70 Hours

Display

Listed below are the modes of operation visible on the display.

Warm-Up

When the power button is pressed, the monitor enters a one

minute warm-up indicated by the word WARM-UP in the upper

left corner.

On Line

Indicates when a PC is communicating to the sensor via the

RJ45 port in the rear of the monitor.

Normal Operating Mode

After warm-up the sensor will stabilize and display current
conditions.

Adjustment modes

Pressing the mode button can access adjustment modes. The

following information is displayed for each mode. Once the

mode is displayed, press enter to adjust the setting. Press enter

again to leave the adjustment mode.

Flashing ELEVATION

 - This feature allows the user to correct

the monitor for elevation changes. ”ELEVATION” will flash in the

lower right corner of the display. Press the enter button and the

current elevation above sea level will be displayed. To change

adjust from ft. to meters press the mode button. Use the up/dwn

buttons to adjust elevation in 500 ft. (100 meter) increments.

When finished with adjustment press enter and the adjustment

will be saved.

Flashing CALIBRATION

 - When the word ”CALIBRATION”

flashes in the middle of the screen the user can enter the CO

2

calibration mode using the procedure described below.

Flashing ”TEMPERATURE”

 - When the word ”TEMPERA-

TURE” is flashing at the bottom of the screen the user can enter

the Temperature calbration mode using the procedure

described below.

Flashing CO

2

 and ppm

When ”CO

2

” and ”ppm” are flashing on the screen the user can

enter mode to adjust the outdoor reference concentration for

the cfm/person calculation.

See the procedure below for more information.

Calibration in Progress Mode

Displays when calibration is in progress.

Outdoor Concentration Mode

To properly calculate the ventilation rate in the space the CO

2

sensor must know what the outside CO

2

 

concentrations are.

This mode allows the CO

2

 sensor to store in it´s memory the

outside concentration.

It is activated by holding the ”enter button” for 5 seconds when

the monitor is exposed to stabilized outside conditions.

When this feature is activated, CO

2

, PPM and the lower display

should flash.

The instructions below also describe how the outside concent-

ration can be manually set.

Low Battery

 - Displays when the power source is low.

Battery Installation - 

Remove the battery cover by pressing

the pressure clip (located on the bottom) and pull upward.

Follow the battery diagram imprinted on plastic forproper

installation.

Replace battery cover when battery installation is complete.

Low Battery 

- A warning signal.

(Low battery)

, will flash when there is less than 30 minutes of

battery life.

At this point the batteries should be replaced or the AC

adapter should be used as a substitute.
If operation continues, the unit will become inoperable and

only the 

Low Battery 

will be blink on the LCD display.

AC Power - 

The sensor is shipped with a 6 Vdc 500 mA AC/

DC adapter.
To use the AC adapter, connect the plug into the back of the

unit and plug the transformer into any standard wall outlet.

NOTE

Use the supplied adapter.

Using the wrong adapter may cause damage to the unit.

If power is lost during operation, battery operation will not

function as a back up.

Reviews: