background image

Control Panel Guide 

LEGEND: 

1 — Digital Display 

6 — External Sensor connector 

1.a Top Display                          7 — Main Power switch. 

1.b Bottom Display                     8 — Power cord attachment. 

2 — Selection button.                         9 —  Fuse holder. 

3 — Increase Button/ Enter button. 

4 — Decrease Button. 

5 — Status LED. 

Suffix guide: 

Set temperature for type 0 and 1 operation. 

Actual Temperature of Sensor 1 (red/blue wire) 

c

 - 

Actual Temperature of Sensor 2 (white wire)

 

D

 - 

Actual Temperature of Internal Sensor

 

C

 - 

Set Temperature for profile.

 

t

 - 

Set duration/time (seconds) for profile.

 

1.a 

Front Panel 

Back Panel

 

1.b

 

OPERATING PROCEDURES 

A.  INITIAL PROCEDURES 

1.  Make sure all switches are deactivated. 

2.  Attach external sensor  probes to the three pin socket. (#6  on 

control panel guide) 

3.  Attach  power  cord  to  the  power  cord  attachment  (#8  on 

control panel guide) 

4.  Plug the device to the main power source. 

5.  To  turn  the  unit  ON.  Toggle  the  main  power  switch    to  ON 

position (#7 on control panel guide)

 

B.  OPERATION TYPE(MODE) SELECTION 

1.  Follow initial procedures, “

A. INITIAL PROCEDURES

”. 

2.  The  display  would  show  “

TYPE  0

”,  which  means  Type  0 

operation will  be used. To  select  between  types  0 to  2.  Press 

the  increase  or  decrease  buttons  (#3  &  #4  of  control  panel 

guide). 

3.  To  confirm  selection  and  enter  into  operation  mode  using the 

selected  type.  Press  the  selection  button.(#2  of  control  panel 

guide)

 

C.  TYPE “0”  OPERATION 

This type  of  operation  utilizes  the  internal  temperature  sensor 

to  control  the  heat.    Using  this  type  frees  up  the  two  extra 

external temperature probes for additional monitoring. Attach the 

two  extra  temperature  probes  to  areas  of  interest  such  as  the 

bottom of the board and top of the board near the components to 

be worked on for monitoring. 

1. 

To  set  the  desired  temperature  press  the  selection  button 

repeatedly  until  the  top  display  shows  “

Set

”.  The  bottom 

display would show the current set temperature followed by 

a suffix “

A

”.

Summary of Contents for INT883

Page 1: ...hows err2 internal temperature connector might have come loose or needs to be replaced If err messages shows err3 the first external temperature probe was not detected secure the external temperature on the underside of the board to be worked on PROBLEM 3 DISPLAY AND OTHER DEVICE OPERATION ISSUES SOLUTION Turn off power and back on OTHER PROBLEMS NOT MENTIONED Contact the vendor IR Preheating Syst...

Page 2: ...3 degrees per second Therefore based on your set time the maximum temperature may not reach 250C Profile analysis SEG1 SEG2 SEG3 SEG4 SEG5 SEG6 050t 060t 080t 050t 040t 60t 100C 150C 180C 190C 195C 100C Segment one is set to 50 seconds to reach 100C Which means after 50seconds at the end of segment one the temperature should reach 100C Then Segment two is set to 60 seconds to reach 150C which mean...

Page 3: ...the profile repeatedly press the selection button until the display shows the word Run Prof then press the increase button A 3 second countdown will commence before automated reworking starts 5 To see the running time or current segment the system is processing or the temperature of the external temperature probes repeatedly press the selection button to switch between different views Follow the s...

Page 4: ...with either hot air or IR top heating systems 9 OPERATING PROCEDURES D TYPE 1 OPERATION Before proceeding with this type of operation attach the first external temperature probe to the underside of the PCB to be worked on The second probe can be placed near areas of interest This type of operation utilizes the first external temperature sensor to control the heat Using this type of operation allow...

Page 5: ...play shows the word Act2 and the bottom display shows the actual temperature of second external temperature probe followed by a suffix c 6 To simultaneously view the actual temperature read by the first and Second external temperature probe repeatedly press the selection button until the top display shows the actual temperature of first external temperature probe followed by a suffix b and the bot...

Page 6: ...ower source 5 To turn the unit ON Toggle the main power switch to ON position 7 on control panel guide B OPERATION TYPE MODE SELECTION 1 Follow initial procedures A INITIAL PROCEDURES 2 The display would show TYPE 0 which means Type 0 operation will be used To select between types 0 to 2 Press the increase or decrease buttons 3 4 of control panel guide 3 To confirm selection and enter into operati...