Anyload OCSL User Manual Download Page 11

 
 

10 

ANYLOAD OCSL Mini Type Crane Scale User’s Guide

 

 

 

 

5.

 

Calibration 

 

Options 

Display 

Operations 

Notes 

Enter 
Calibration 
Mode 

 

Hold and press [HOLD] and 
[TARE] simultaneously, keep 
pressing until SETUP 
displays. 

Enter Setup Mode 

(At this stage set the unit of 
measurement you are going to 
calibrate, hold & press 
the[HOLD/UNIT] to switch unit) 

SETUP 

Press [ON/OFF], [HOLD] and      
[TARE] simultaneously, keep 
pressing until SCALE 
displays. 

Enter Calibration 
Mode 

SCALE 

Press [TARE]. 

Full Scale 

_XXXX 

Press  [HOLD]  to  change 
Full  Scale  and  press [TARE] 
to save setting. 

For example, if you set 
the unit at kg and setting 
the parameter here to 
600 then the full scale 
should be 600kg else if it 
was set to lb then the full 
scale shall be equivalent 
to 600lb. 

Resolution 

E   X 

Press  [HOLD]  to  change 
Resolution  and  press 
[TARE] to save setting. 

1/2/5 

Decimal 
Point 

Pt  X 

Press [HOLD] to change 
Decimal Point and press 
[TARE] to save setting. 

0: XXXXX 
1: XXXX.X 
2: XXX.XX 
3: XX.XXX 

Zero 

Calibration 

 

LoAd0 

Press [TARE] to start zero 
calibration. 

Set the scale to no load 
status and must be at stable 
state. 

   ----- 

Load1 

Calibration 

 

_XXXX 

Press [HOLD] to change 1st   
Calibration Weight value. 
Load the actual weight then 
press [TARE]   to start span 
calibration. 

If using test weight 
equivalent to full scale 
(FULL), the calibration shall 
be finished at this stage. 

 

----- 

Load2 

Calibration 

 

_FULL 

Load the 2

nd

 test weight 

then press [TARE] to finish 
the calibration. 

If 2

nd

 test weight is not 

needed, you can skip this 
stage by pressing the 
[ON/OFF] to finish and exit 
the Calibration. 

 

----- 

 

_End_ 

 

 

 

Summary of Contents for OCSL

Page 1: ...OCSL Mini Type Crane Scale User s Guide...

Page 2: ...1 ANYLOAD OCSL Mini Type Crane Scale User s Guide...

Page 3: ...F CONTENTS 1 Introduction 3 2 Specifications 3 Features 3 Specifications 4 Capacities Resolution 4 Schematic Diagram 5 Dimensions 5 3 Operation Guide 6 Power On 6 Power Off 6 Tare 6 Hold 7 Unit Switch...

Page 4: ...ting the scale this may damage the load cell o Do NOT use scale under thunder or rain o Do NOT attempt to repair the scale by yourself Contact your local dealer or to the Technical Support 2 Specifica...

Page 5: ...activities within 3 min Overload 100 F S 9e Max Safety Load 120 F S Ultimate Load 300 F S Battery Life 150 hours Temp Op 10 40 Humidity Op 90 at 20 Display 0 7 inch 17 78mm numerical Net Weight 620g...

Page 6: ...5 ANYLOAD OCSL Mini Type Crane Scale User s Guide SCHEMATIC DIAGRAM DIMENSIONS...

Page 7: ...Display will show power off message TARE Press to store the weight as tare o TARE will appear and weight reading will turn to zero o If weight is greater than 100 F S or lesser than 0 taring is not a...

Page 8: ...hide and weight reading will resume UNIT SWITCH Press for 1 second to switch measuring unit between kg lb and N o If scale is in tare or in hold unit switching is not allowed ZERO Press for 1 second...

Page 9: ...bled 2 2 F S 4 4 F S 10 10 F S 20 20 F S 50 50 F S Press to enter Manual Zero options and press to select Manual Zero option o Manual Zero can also be set to 0 disabled 2 2 F S 4 4 F S 10 10 F S 20 20...

Page 10: ...FF key for longer period no action taken after TARE or HOLD is pressed defective TARE or HOLD key clean TARE or HOLD key unstable readings scale in motion stabilize the load and scale scale is damped...

Page 11: ...if it was set to lb then the full scale shall be equivalent to 600lb Resolution E X Press HOLD to change Resolution and press TARE to save setting 1 2 5 Decimal Point Pt X Press HOLD to change Decima...

Page 12: ...ANYLOAD Weigh Measure Inc Website www anyload com Email info anyload com Fax 1 866 612 9088 North America Toll Free 1 855 ANYLOAD 269 5623...

Reviews: