Rinsing the Ultra Filter Cartridge

The Ultra Filter Cartridge must be rinsed before the system is 

used to remove any air and protective solution.

NOTE:

 Ultra Filter Cartridge must be rinsed to drain 

before use. Rinsing to drain removes storage solu-

tion and air. Do NOT rinse into carbon if present. 

Carbon life and/or performance may be affected.

NOTE:

 Make sure the Carbon Block Cartridge is NOT 

installed in the Carbon Housing.

1. 

Direct water from the Permeate Outlet to a drain.

2. 

Open the tap or faucet closest downstream to the filter 

system.

3. 

Slowly open the inlet water valve and allow water to enter 

the system.

4. 

Press the 

FLUSH

 button. The Drain Valve will open and 

air and water will come out of the drain line. Repeat six 

times to ensure all air and storage solution are flushed out 

of the Ultra Filter Cartridge.

5. 

Allow water to continue to run through the system and out 

the faucet for 15 minutes.

6. 

Inspect the unit for any leaks and repair as  

needed.

7. 

After 15 minutes with water flow, close the tap or faucet 

and let the system stand without water flow for 15 minutes 

to allow trapped air to come out of the hollow fibers of the 

Ultra Filter Cartridge.

8. 

After 15 minutes without water flow, open the tap or faucet 

for five minutes to flush out the trapped air.

9. 

Close the tap or faucet.

10. 

Press the 

FLUSH

 button to open the Drain Valve. Repeat 

six times to ensure any remaining air is flushed out of the 

system.

Sanitizing the System and Lines

The plumbing must be sanitized to eliminate possible contami-

nation that may have occurred during the installation process. 

One ounce (30 ml) of liquid chlorine bleach (regular bleach, 

unscented 6% sodium hypochlorite) can be used to sanitize 

the plumbing. The amount of bleach needed depends on 

the system and the amount of plumbing downstream of the 

system. 

NOTE:

 Follow the handling and safety instructions  

supplied with the bleach.

1. 

Follow the steps in the 

Rinsing the Ultra Filter Cartridge

 in 

the Installation section of this manual. 

2. 

Turn off the water to the system.

3. 

Open the faucet or tap closest downstream to the  

system. 

4. 

Close the Inlet Water Valve and allow the system to 

depressurize.

5. 

Place a bucket under the “T” or cross fitting at the inlet con

-

nection to the system. Open the inlet drain valve or unscrew 

the cap to drain the system.

6. 

Press the 

FLUSH 

button to help drain the system.

7. 

When the water flow out of the inlet fitting stops, close the 

inlet drain valve and pour the liquid bleach into the inlet 

connection fitting. Be careful not to spill bleach onto cloth

-

ing or skin. You may want to add the bleach using a cup. 

Reattach the cap on the fitting.

8. 

Slowly open the inlet water valve and allow water to flow 

out of the tap until the smell of bleach is present.

9. 

Immediately close the tap and let the system stand with 

no water flow for at least 15 minutes to allow the bleach to 

sanitize the pipes.

10. 

After 15 minutes without water flow, open the tap and 

flush until the presence of bleach is gone. All other taps 

should be opened to flush any bleach from the plumbing. 

Close the taps.

11. 

Press the 

FLUSH

 button to complete the flushing pro

-

cess.

Summary of Contents for 9700960

Page 1: ... P N 1011532 Rev B 04 20 UF 216CC Owner s Manual Water Filtration System Manufacturing Numbers 9700960 www antunes com CAUTION Read all instructions before using the unit Original Instructions ...

Page 2: ...icular application All filter configurations utilize NeoH capillary membranes providing the latest innovation in reusable surface filtration technology This manual provides the safety installation and operating procedures for the UF 216CC Series water filtration system We recommend that all information contained in this manual be read prior to installing and operating the unit Your UF 216CC Series...

Page 3: ...lugging the power cord Water pressure must not exceed 100 psig 690 kPa To reduce water pressure install a water pressure regulator and set water pressure to suit application The trans membrane pressure inlet water pressure minus the permeate water pressure must not exceed 45 psi 310 kPa System should be supplied only with cold water Commonwealth of Massachusetts Plumbing Code 248 CMR shall be adhe...

Page 4: ...B Height C UF 216CC 9700960 16 3 41 4 cm 6 8 17 3 cm 24 2 61 5 cm Electrical Cord Plug Configurations Kit Model Number Description Configuration 0012146 DC Power Supply 100 240 VAC Includes the 4 plug adaptors below NOTE PLUG STYLE MAY VARY US NEMA 1 15 2 pin or NEMA 5 15 3 Pin US NEMA 1 15 2 PIN or NEMA 5 15 3 PIN Euro CEE 7 16 EURO CEE 7 16 AS NZS 3112 AUS 2 Pin AS NZS 3112 2 PIN AUS UK BS 1363 ...

Page 5: ... power outage or the system is unplugged water will continue to be filtered but the system will NOT flush This could cause the Ultra Filter to plug prematurely and may reduce the life of the filter The flush is automated by the controller to last 10 seconds and take place in one hour intervals This interval can be changed see Changing the Interval Setting in the Maintenance section of this manual ...

Page 6: ... highly recommended for ease of installation and future servicing CAUTION This equipment is to be installed to comply with the basic plumbing code of the Building Officials and Code Administra tors Inc BOCA and the Food Service Sanitation Manual of the Food and Drug Administration FDA CAUTION Water pressure must not exceed 100 psig 690 kPa To re duce water pressure install a water pressure regulat...

Page 7: ...ate connection This will help prevent possible contamination of the filter system due to other equipment downstream The check valve not supplied should be mounted close to the system outlet and sized properly for the plumbing line Check with local codes for the proper specification A shutoff valve not supplied should be installed in the filtered water line leading from the system The valve should ...

Page 8: ...system Sanitizing the System and Lines The plumbing must be sanitized to eliminate possible contami nation that may have occurred during the installation process One ounce 30 ml of liquid chlorine bleach regular bleach unscented 6 sodium hypochlorite can be used to sanitize the plumbing The amount of bleach needed depends on the system and the amount of plumbing downstream of the system NOTE Follo...

Page 9: ...d stop at the end of the thread Use a Filter Wrench if necessary NOTE Do NOT overtighten the Carbon Element Housing 3 Direct water from the Permeate Outlet to a drain 4 Turn on the water to the system by slowly opening the inlet water valve 5 Allow water to flow through the Carbon Element and Carbon Element Housing to flush out trapped air and any debris that may be present 6 To help remove trappe...

Page 10: ...ope attach the elbow fitting to the new cartridge and the inlet strainer to the inlet end 15 Insert the new cartridge into the housing 16 Using thread seal tape and pipe dope attach the elbow and drain solenoid valve to the drain end of the cartridge NOTE Do NOT overtighten the elbows on to the cartridge 17 Using a screwdriver tighten the cartridge clamp screws to attach the cartridge to the housi...

Page 11: ...oved food grade lubricant and install into the Carbon Element Housing 11 Install the new Carbon Element into the Carbon Element Housing NOTE Make sure the Gasket remains in place in the recessed grooves on both ends of the Carbon Element 12 Screw the Carbon Element Housing containing the Carbon Element into place on the Carbon Element Head The Housing will come to a hard stop at the end of the thr...

Page 12: ...restored based on the interval setting selected Figure 6 Controller Flush Interval Button Hold Time Button LED Flush LED 15 minutes A 10 seconds On On 30 minutes B 10 seconds On On 45 minutes C 10 seconds On On 1 hour A 5 seconds On Off 4 hour B 5 seconds On Off 6 hours C 5 seconds On Off 12 hours D 5 seconds On Off 24 hours D 10 seconds On On Caution Changing the flush interval can cause the Ultr...

Page 13: ...tion not completely rinsed out of system Rinse system for a longer period of time Biological growth in pipes Sanitize plumbing Water conditions changed Consider installing additional filtration Broken capillary in Filter Cartridge Replace Filter Cartridge Flush runs continuously Drain Valve stuck open Replace rebuild the Drain Valve Controller sending continuous signal to valve Replace the control...

Page 14: ... 4MPT X 1 2MPT 2 Replacement Parts Note Use only genuine Antunes replacement parts in this unit Use of parts other than those supplied by the manufacturer will void the warranty Replacement Parts can be purchased from an authorized dealer Contact us at 86 512 6841 3637 or toll free in the United States at 1 800 253 2991 Item Part No Description Qty 9 2190194 Elbow 90 1 2 NPT Reinforced 2 10 219020...

Page 15: ...UF 216CC WATER FILTRATION SYSTEM 15 P N 1011532 Rev B 04 20 Notes ...

Page 16: ...nce calibration or regular adjustments as specified in operating and maintenance instructions of this manual and or labor involved in moving adjacent objects to gain access to the Equip ment 2 Antunes reserves the right to make changes in design or add any improvements on any product The right is always reserved to modify equipment because of factors beyond our control and government regulations C...

Reviews: