Maintenance
11.1 Changing the Batteries
You can see the status of the battery by looking at the battery symbol in the top right-
hand corner of the display.
The following status are possible:
When the batteries are almost empty, the warning "Low Battery " is displayed. After
the second warning "Battery Empty", the instrument switches itself off.
If the batteries are flat, you can easily replace them.
1.
Place the filling tube in a vessel which is suitable for sample disposal.
Battery is full
Battery is a little low
Battery is low
Battery is empty
WARNING
When opening the battery compartment or exchanging batteries,
sparks may be generated which can cause an explosion or fire in
hazardous areas. Serious injuries are possible.
•
Never open the battery cover in hazardous areas.
•
Only replace the batteries outside hazardous areas.
NOTICE
•
Always observe the correct polarity when replacing the
batteries.
•
Only use batteries of the same type and with the same level of
battery charge.
•
For DMA 35 instruments with ATEX marking (DMA 35 Ex and
DMA 35 Ex Petrol), only use the permitted alkaline battery type
EN91 (LR06, AA) from Energizer Industrial.
Summary of Contents for 84138
Page 2: ......
Page 3: ...C96IB01 3 EN DMA 35 Portable Density Meter From Firmware Version V1 79 Instruction Manual ...
Page 72: ...72 C96IB01 EN Appendix C CE Declarations of Conformity ...
Page 73: ...C96IB01 73 EN ...
Page 74: ...74 C96IB01 EN Appendix D EC Type Examination Certificate ...
Page 75: ...C96IB01 75 EN ...
Page 76: ...76 C96IB01 EN Appendix E RFID Module Test Reports ...
Page 77: ...C96IB01 77 EN ...
Page 81: ...C96IB01 81 DE DMA 35 Tragbares Dichtemessgerät Ab Firmwareversion V1 79 Betriebsanleitung ...
Page 153: ...C96IB01 153 DE Anhang C CE Konformitätserklärung ...
Page 154: ...154 C96IB01 DE ...
Page 155: ...C96IB01 155 DE Anhang D EG Baumusterprüfbescheinung ...
Page 156: ...156 C96IB01 DE ...
Page 157: ...C96IB01 157 DE Anhang E RFID Modul Testreports ...
Page 158: ...158 C96IB01 DE ...