Anthogyr SAFE Instructions For Use Manual Download Page 4

22 

Removing a tool: 

 Align the arrow on the ferrule with the open padlock 

 Withdraw the tool by pulling on it 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

  After insertion of an instrument, check its correct retention 

  with a slight axial movement. 

It is essential that the motor is stopped and outside of the patient's 

mouth whenever the instrument is inserted or extracted. 
The MD (Im Instruments) does not require maintenance 

work. Nevertheless, should you notice a functional defect in the 

locking or gripping mechanism, this could  mean that it is dam-

aged. In this case, immediately return the device to the Anthogyr 

after-sales service department. 

Make sure that the device does not have any areas of corrosion or 

cracks, and check that it is working properly. 

MD

 lifecycle  

If used in a proper manner, all 

MD

 parts have a lifecycle corre-

sponding to 250 sterilisation cycles. 

However, these indications are not a warranty because wear may 

appear  prematurely,  depending  on  how  the 

MD

  is  maintained 

(cleaning and sterilisation). 

 

VII. HYGIENE AND MAINTENANCE 

  Sterilisation of MD (Im Instruments)  must be per

  formed  by  properly  trained  and  protected  personnel,  in 

accordance  with  the  applicable  regulations.  The  sterilisation 

protocol must be adapted for the risk of infection. 

  In order to avoid any risk of infection and injury, it is essen-

  tial to wear protective gloves. 
  Refer to the manufacturer’s instructions for each product 

  used. Observe the concentrations, lengths of exposure and 

the  lifespan  of  the  products.  Do  not  combine  products  and  ob-

serve the provisions regarding their disposal. 

  Only use products that are intended for the maintenance of 

  medical and surgical equipment, compatible with stainless 

steel  (no  chlorine  compounds),  and  compatible  with  aluminium 

(no soda). 
Avoid the use of antiseptics that are intended for use on the skin 

and on mucous membranes. 

Avoid the use of products containing aldehydes, alcohol or other 

products that are likely to bind proteins. 

Preference should be given to detergents / disinfectants that are 

pH neutral or slightly alkaline. 
The device must be cleaned and sterilised after every procedure. 
The motor must be disconnected or stopped before handling. 
Instruments must not be connected to the MD Impactor. 
The MD Impactor is not removable and is not submersible. 
The instruments are submersible. 

7.1. Manual preparation for sterilisation: 

Reprocessing of the MD (Im Instruments) must be carried 

F

ra

n

ça

is

 

En

g

li

sh 

(U

K

)

 

D

e

u

tsch

 

Esp

a

ñ

o

It

al

iano

 

P

o

rt

u

g

u

ê

s

 

N

e

d

e

rla

n

d

s

 

中国的

 

23 

   

  ر

ع

 

En

g

li

sh 

(U

S

)

 

out as soon as possible after treatment (within max. 2 hours). 
By brushing: 

 Brush under running water using a soft brush. 

 Clean the entire MD (Im Instruments) using a disinfectant 

wipe. 
Or by spraying: 

 Spray the disinfectant on to the MD (Im Instruments) and 

wipe using a clean cloth. 
Rinsing and drying 

 Rinse properly

 1

 

 Dry the entire MD (Im Instruments). 

7.2. Automatic preparation for sterilisation: 

  Only  use  a  washer-disinfector  appropriate  for  the  repro-

  cessing of this type of MD (Im Instruments). Follow 

the instructions provided by the manufacturer of the unit. 

 If the unit is equipped with nozzles for contra-angles: Fit the MD 

Impactor onto the nozzle of the device. 

 If the unit is not equipped with nozzles for contra-angles: without 

disassembling the MD impacor, position it with the tool connection 

component pointing upwards. Position the parts so that any reten-

tion of fluid is avoided. Immobilise each part. 

 The thermal disinfection cycle must be at least 10 minutes at 93°

C (203°F). 

 Follow the instructions for use provided by the thermal disinfector 

manufacturer,  and  in  particular  the  recommendations  regarding 

the products that should or should not be used with the unit. 

 Observe the drying cycle (Caution: Do not exceed 140°C (284°F) 

during the drying cycle) 

  Check  that  there  is  no  residue  and  that  all  parts  of  the  MD 

(Im Instruments) are perfectly dry at the end of the cycle. 
The MD impactor is not submersible. 

7.3. Lubrication: 

  

 Wear a protective face mask. 

 

 Disconnect the MD (Impactor) from the motor before handling. 

 Remove the connected instrument in the MD (Impactor) 

 Check that the nozzle of the spray is the correct one for use with 

the MD (Impactor). 

 Insert the aerosol spray at the back of the MD (Impactor). 

 Cover the head (connecting part) of the MD (Impactor) with a wipe. 

 Position the head (connecting part) of the MD (Impactor) down-

ward. 

 Spray several times, keeping the MD (Impactor) closed. 

 Wipe off surplus oil with a soft cloth or a wipe. 

  Lubricate  after  each  decontamination  and  before  each 

  sterilisation, and repeat in the event of prolonged use. 

  Keep  away  from  heat  and  ignition  sources.  Smoking  is 

  strictly prohibited! 

7.4. Functional test: 

  This test must be performed every time prior to sterilisation. 

  Connect the MD (Impactor) to the motor and point the head 

downwards. Run at low speed (5000 motor rpm) for 30 seconds in 

order to clear the excess oil, then at full speed (10,000 motor rpm) 

for 30 seconds. Wipe the MD (Impactor) with a disinfectant wipe if 

the oil has dripped. 
Note: If you notice overheating, irregularities, vibrations or abnor-

mal  sounds  during  operation  of  the  MD  (Impactor)  ,  immediately 

contact your authorised technical service centre. 

7.5. Sterilisation: 

  Resterilisation  of  reusable  MDs  must  be  performed  by 

  properly  trained  and  protected  personnel,  in  accordance 

with  the  applicable  regulations.  The  resterilisation  protocol  must 

be adapted for a risk of infection. 

  The MD (Im Instruments) and the accessories must 

  be  sterilised  before  they  are  used  for  the  first  time,  and 

every time after they are used (after cleaning and disinfection). 

Do not sterilise instruments unless they have been pre-disinfected, 

cleaned,  decontaminated,  lubricated  (only  the  MD  Impactor)  and 

tested. 

  No disassembly of the MD is envisaged prior to performing 

  sterilisation. 

: Temperature <38°C, demineralised water is recommended instead of running water, for example 

if the latter contains too much chlorine (see the FD98-135 standard)  

Reviews: