Anthogyr SAFE Instructions For Use Manual Download Page 3

20 

VI. MD START UP PROCEDURE  

 

The MD (Im Instruments) are supplied non-sterile and 

  non-lubricated*.

 

Before  using  for  the  first  time,  the  MD  (Impactor  +  Instruments) 

must be cleaned, decontaminated, lubricated* and sterilised  (see 

the section on "Hygiene and maintenance"). 

*: for the “Impactor SAFE3”

 

MD only 

  Resterilisation  of  reusable  MD  (Impactor  +  Instruments) 

  must  be  performed  by  properly  trained  and  protected 

personnel,  in  accordance  with  the  applicable  regulations.  The 

resterilisation protocol must be adapted for a risk of infection. 

  Before  use,  check  that  the  MD  (Impactor  +  Instruments) 

  and/or the osteotomes are not damaged and that no parts 

are  missing.  Check  the  condition  of  the  instruments  used  and 

handle  them  with  caution  and  care. 

(The  list  of  accessories  is 

included in paragraph IX) 

6.1. Motor connection:

 

  Protective gloves should preferably  be used whenever  the 

  tools are handled. 

Check the condition of the MD Impactor used and handle it with due 

care and attention. 

  The  MD  (Impactor)  may  only  be  connected  to  the  motor 

  outside of the patient's mouth and when the motor has been 

stopped. 
Position the MD (Impactor) on the motor coupling (in accordance 

with EN 23964 - ISO 3964) until it clicks into place. Keep the motor 

and the MD (Impactor) on the same axis for this purpose. 
 
 
 
 
 
 

 

For  certain  motors,  it  is  necessary  to  activate  the  side  button  to 

facilitate the connection. 

Pull gently on the MD (Impactor) every time before use in order to 

verify that it is correctly engaged on the motor coupling. 
Functional test 

  Run the motor in the normal  direction (the device will  not     

  operate if the motor is in "reverse" mode). Check that the 

MD (Impactor) properly generates regular micro-impacts. 

NOTE: If you notice overheating, irregularities, vibrations or abnor-

mal  sounds  during  operation  of  the  MD  (Impactor),  immediately 

contact your authorised technical service centre. 

6.2. Disconnecting the motor: 

  CAUTION:  The  MD  (Impactor)  may  only  be  disconnected 

  when  the  motor  has  been  stopped,  and  outside  of  the 

patient's mouth. Withdraw the MD (Impactor) while maintaining the 

axis of the motor. For certain motors, it is necessary to activate the 

side button to enable the MD (Impactor) to be disconnected. 
NOTE: To prevent oil from leaking into the motor, do not leave the 

MD (Impactor) connected to the motor if the MD (Impactor) is not 

used for prolonged periods. 

 
 
 
 
 
 
6.3. Connecting / disconnecting the instrument: 

  The instruments may only be connected to the MD 

Impactor 

  when  the  motor  is

  switched  off  and  outside  the  patient’s 

mouth. 

  Protective gloves should preferably  be used whenever  the 

  tools are handled. 

  Check  the  condition  of  the  instruments  used  and  handle 

  them with caution and care. 

F

ra

n

ça

is

 

En

g

li

sh 

(U

K

)

 

D

e

u

tsch

 

Esp

a

ñ

o

It

al

iano

 

P

o

rt

u

g

u

ê

s

 

N

e

d

e

rla

n

d

s

 

中国的

 

21 

   

  ر

ع

 

En

g

li

sh 

(U

S

)

 

6.4. Locking/unlocking the 

MD

 (Instruments) in the 

MD

 

(Impactor) : 

Positioning a tool: 

 Align the arrow on the ferrule with the open padlock 

 

 

 

 

 

 

 

 

 

 

 

 

 If the MD (Instruments) has a trefoil symbol, align the symbol on 

the MD (Instruments) with the symbol at the end of the MD  Im-

pactor connection and insert the MD (Instruments). 
 
 
 
 
 
 

 
 

 If the MD  (Impactor) is standard, insert the tool by locating the 

drive bar. 

• Turn the collet to align the arrow with the closed padlock and 

lock the MD (Impactor) on the MD (Impactor).  
 

 

 

 

 

 

 

 

 

 

 

 

 Check that the MDs (Instruments) are firmly in place by applying 

gentle traction to them. 
 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 