5-11 High Band SPDT Switch Control PCB Assembly
Chapter 5 — Component Replacement
5-14
PN: 10410-00730 Rev. D
MN469xC Series Multiport Test Set MM
5-11 High Band SPDT Switch Control PCB Assembly
Assembly: ND70926<R> or 3-ND70926<R>
This section provides a procedure for removing and replacing the High Band SPDT Switch Control PCB
Assembly. The Switch Control PCB Assembly is mounted on top of the High Band SPDT Switch Assemblies
A12, A13, A18, and A19; Parts 70241 or 70242.
A12, A13, A18 or A19 Replacement
1.
Remove the top cover from the test set as instructed in
2.
Replace the Switch Control PCB as illustrated in
Caution
Do not force the Control PCB onto the Switch Assembly. The Control PCB must be in good
alignment with the Switch contact pins.
1. High Band Switch Control Board locations
2. Control Board Removal:
Disconnect the switch control cable from the High Band SPDT Switch Control PCB Assembly.
Remove the three mounting screws from the Switch Control PCB. (Save the screws and washers).
Unplug the Switch Control PCB from the High Band SPDT Switch.
3. Installation is the reverse of removal.
Figure 5-13.
High Band SPDT Switch Control PCB Assembly,
1
2
ND70926
High Band
Switch Control Board