![Anritsu MU181800A Operation Manual Download Page 38](http://html.mh-extra.com/html/anritsu/mu181800a/mu181800a_operation-manual_2945430038.webp)
Chapter 6 Performance Test
6-4
6.3 Performance Test Items
This section describes the following test items.
•
Levels at Clock Input and Clock Output
6.3.1 Measuring levels at Clock Input and Clock Output
(1) Specifications
Table 6.3.1-1 Specifications for Clock Input level and Clock Output level
Clock Input level
0.4 to 2.0 Vp-p
Clock Output level
0.4 to 1.0 Vp-p (Fixed)
Test condition
Clock frequency: 0.1 to 12.5 GHz (MU181800A)
0.1 to 14.0 GHz (MU181800B)
0.1 to 14.1 GHz (MU181800B-005)
(2) Connection
MP1800A
MU181800A
Clock Output #1
MU181020A
External Clock Input
AUX out
Clock Outputs
#2 to #4
Synthesizer
Clock Input
Trigger
Electrical input
Sampling Oscilloscope
Fig. 6.3.1-1 Connection diagram for Clock Input level and Clock Output level measurement