![Anritsu MU181800A Operation Manual Download Page 19](http://html.mh-extra.com/html/anritsu/mu181800a/mu181800a_operation-manual_2945430019.webp)
1.3 Specifications
1-5
1.3 Specifications
Table 1.3.1-1 Specifications
Item Specifications
Remarks
Clock input
Operating
frequency
0.1 to 12.5 GHz (MU181800A)
0.1 to 14.0 GHz (MU181800B)
0.1 to 14.1 GHz (MU181800B-005)
Level
0.4 to 2.0 Vp-p
Waveform
0.5 GHz square wave
>0.5 GHz square or sine wave
Connector SMA
Termination 50
/GND
Clock output
Channel
Single-end, 4 systems (MU181800A)
Single-end, 5 systems (MU181800B)
Interchannel
skew
10 ps (at 12.5 GHz, MU181800A)
10 ps (at 14.0 GHz, MU181800B)
10 ps (at 14.1 GHz, MU181800B-005)
Level
MU181800A: Min. 0.4 Vp-p
,
Max. 1.0 Vp-p
MU181800B: Min. 0.4 Vp-p
,
Max. 1.2 Vp-p
(
1 GHz)
Min. 0.4 Vp-p, Max. 1.0 Vp-p
(>1 GHz)
Duty 50
10% at Clock input duty 50%
Connector SMA
Termination 50
/GND
Mechanical
performance
Dimensions
234 mm (W)
21 mm (H)
175 mm (D)
(Compact-PCI 1 slot) (Protrusion excluded)
Mass
1.5 kg
Environmental
performance
Storage
temperature
range
20 to +60
C (recommended range: +5 to +30
C)
Storage humidity
range
20 to 80% (recommended range: 40 to 75%)
Channel1
Channel2
Channel4 or 5
...
Summary of Contents for MU181800A
Page 10: ...x ...
Page 14: ...IV ...
Page 20: ...Chapter 1 Overview 1 6 ...
Page 24: ...Chapter 2 Preparation before Use 2 4 ...
Page 30: ...Chapter 3 Panel Layout and Connectors 3 6 ...
Page 40: ...Chapter 6 Performance Test 6 6 ...
Page 47: ...Appendix App I Appendix A Performance Test Result Sheet A 1 ...
Page 48: ...Appendix App II ...
Page 50: ...Appendix A Performance Test Result Sheet A 2 ...