Section 3 Operation
3-30
3.7
Coupled Function
3.7.1
Resolution Bandwidth (RBW) and Sweep Time
To set the RBW and Sweep Time, perform the following key operations.
RBW
Manual
RB, VB, SWT
Auto
All Auto
Auto
Sweep
Time
Manual
RB, VB, SWT
Auto
All Auto
Auto
Manually set the RBW with the ten keys, step keys,
and rotary knob.
Automatically set the RBW.
Automatically set the RBW, VBW, and Sweep Time.
Automatically set the RBW, VBW, Sweep Time, and Atten.
Manually set the Sweep Time with the ten keys, step keys,
and rotary knob.
Automatically set the Sweep Time. (Note)
Automatically set the RBW, VBW, and Sweep Time.
Automatically set the RBW, VBW, Sweep Time, and Atten.
(1)
Auto mode
The RBW. Sweep Time, and VBW parameters are set to Auto so that even if the
frequency span is varied, the respective parameters are automatically set to the
optimum values so that frequency and level measurement errors do not occur.
The following shows the Swp Time Auto setting range:
• Lower limit value
100 msec
• Upper limit value
1000 sec