Remote Operation Command Details
Remote Operation
4-4
PN: 10585-00016, Rev. A
MA24104A UG
HAT
Description:
Sets the high aperture time mode.
Syntax:
HAT + LF
Return Value:
OK or ERR
Remarks:
This command will put the sensor in high aperture time mode. In this mode, the A to D
converter integration time is about 160 milliseconds.
LAT
Description:
Sets the low aperture time mode.
Syntax:
HAT + LF
Return Value:
OK or ERR
Remarks:
This command will put the sensor in low aperture time mode. In this mode, the A to D
converter integration time is about 10 milliseconds. This mode is the default mode for the
sensor when powered up.
ASDON
Description:
Turns on auto shutdown mode
Syntax:
ASDON + LF
Return Value:
OK or ERR
Remarks:
This command applies when RS232 connectivity is used with battery power. This is the
default mode at power up. In this mode, the sensor will automatically go into sleep mode
to conserve battery power if the RS-232 serial port cable is disconnected from the host.
The sensor will wake up when an active host is reconnected and must be zeroed before
taking measurements.
ASDOFF
Description:
Turns off auto shutdown mode
Syntax:
LF
Return Value:
OK or ERR
Remarks:
This command applies when RS232 connectivity is used with battery power. In this
mode, the sensor will not go into sleep mode if the RS-232 serial port cable is
disconnected from the host.