Channel Scanner (Option 27)
4-6 Script Master Measurement Setup
Spectrum Analyzer MG
PN: 10580-00349 Rev. M
4-5
4.
Press the
Repeat Scan Type
to select the desired scan mode
# Scans
, go to step 4a, to
select
Time
, go to step 4b.
a.
If # Scans is selected, press # of Repeats (List) to set the desired repetition of test
cycles of the Script Master Test File. Select the # of Repeats (List) define how
many channels to include in the custom list. This choice can be changed later if
needed.
b.
If Time is selected, press Scan Duration to set the desired testing period. Time on
the submenu key turns red for editing. Pressing a number on the numeric keypad
lists the Time menu. Press the appropriate time unit.
5.
If repeated set testing is desired, press the
# of Repeats (Sets)
submenu key and enter
the desired number of test cycles.
6.
Press the
Record to the On position
to store test measurements.
7.
Press the
Start/Restart Test
submenu key to begin testing.