Check the Bag Suspender
The bag suspender should be checked every three to six months or if fiber values are higher than normal or
inconsistent. To check the Bag Suspender, follow the steps below.
1.
Verify that you are using the
NEW
version of the Bag
Suspender Trays.
The NEW version of the trays has 9 holes in groups of 3
as shown in the picture to the right.
NEW
OLD
You CANNOT use the new and old bag suspender trays together.
2.
Check the trays for melting.
The pictures shown are examples of extreme cases.
However, for proper operation you must replace trays
that show signs of melting or wear.
3.
Check that the bottom tray is flat.
If the bottom tray is concave (see picture) the bag
suspender will catch in the vessel and melt.
Troubleshooting & Replacement Parts
The ANKOM Technology web site has the most current troubleshooting and replacement parts information.
Therefore, if you have any questions about the operation of your ANKOM
2000
Fiber Analyzer, or if you need
replacement parts, please visit our web site at
www.ankom.com.
Melted trays
Concave
Tray
Summary of Contents for 2000
Page 2: ...This page intentionally left blank ...
Page 4: ...This page intentionally left blank ...
Page 18: ...Operator s Manual pg 18 Rev 05 12 21 This page intentionally left blank ...
Page 24: ...Operator s Manual pg 24 Rev 05 12 21 This page intentionally left blank ...
Page 30: ...Operator s Manual pg 30 Rev 05 12 21 This page intentionally left blank ...
Page 40: ...Operator s Manual pg 40 Rev 05 12 21 Appendix B Wiring Diagrams 1 of 4 ...
Page 41: ...Operator s Manual Rev 05 12 21 pg 41 Appendix A Wiring Diagrams 2 of 4 ...
Page 42: ...Operator s Manual pg 42 Rev 05 12 21 Appendix A Wiring Diagrams 3 of 4 ...
Page 43: ...Operator s Manual Rev 05 12 21 pg 43 Appendix A Wiring Diagrams 4 of 4 ...