  DN:KD-322-SMSY01 V4.0 

Operation Guide 

Template version:1

 

1/14 

MODEL KD-322

 

Semi Automatic Arm Cuff Blood Pressure Monitor 

  (ELECTRONIC SPHYGMOMANOMETER) 

OPERATION GUIDE 

__________________________________________________________________________ 

INDEX 

 

IMPORTANT INFORMATION ......................................................................... 2

 

CONTENTS AND DISPLAY INDICATORS ..................................................... 2

 

INTENDED USE .............................................................................................. 3

 

CONTRAINDICATION ..................................................................................... 3

 

PRODUCT DESCRIPTION .............................................................................. 3

 

SPECIFICATIONS ........................................................................................... 4

 

NOTICE ........................................................................................................... 4

 

SETUP AND OPERATING PROCEDURES .................................................... 6

 

1.

 

BATTERY LOADING ............................................................................... 6

 

2.

 

CONNECTING THE CUFF TO THE MONITOR ....................................... 6

 

3.

 

APPLYING THE CUFF ............................................................................. 6

 

4.

 

BODY POSTURE DURING MEASUREMENT ......................................... 7

 

5.

 

TAKING YOUR BLOOD PRESSURE READING ..................................... 7

 

6.

 

ASSESSING HIGH BLOOD PRESSURE FOR ADULTS ........................ 8

 

7.

 

TROUBLESHOOTING (1) ........................................................................ 9

 

8.

 

TROUBLESHOOTING (2) ........................................................................ 9

 

MAINTENANCE ............................................................................................ 10

 

EXPLANATION OF SYMBOLS ON UNIT ..................................................... 10

 

WARRANTY INFORMATION ........................................................................ 11

 

SERVICE CENTER ....................................................................................... 11

 

ELECTROMAGNETIC COMPATIBILITY INFORMATION ............................ 11

 

 

Summary of Contents for KD-322

Page 1: ...RS 2 INTENDED USE 3 CONTRAINDICATION 3 PRODUCT DESCRIPTION 3 SPECIFICATIONS 4 NOTICE 4 SETUP AND OPERATING PROCEDURES 6 1 BATTERY LOADING 6 2 CONNECTING THE CUFF TO THE MONITOR 6 3 APPLYING THE CUFF 6...

Page 2: ...The highest value usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while most peop...

Page 3: ...egrated pressure sensor blood pressure and pulse rate can be measured automatically and non invasively The LCD display will show blood pressure and pulse rate The Electronic Blood Pressure Monitor cor...

Page 4: ...on 90 RH 13 Environmental temperature for storage and transport 20 55 4 131 14 Environmental humidity for storage and transport 90 RH 15 Environmental pressure 80kPa 105kPa 16 Battery life Approx 540...

Page 5: ...ted that you consult with your physician for accurate assessment 13 Please do not use the cuff other than supplied by the manufacturer otherwise it may bring biocompatible hazard and might result in m...

Page 6: ...e 2 CONNECTING THE CUFF TO THE MONITOR Insert the cuff tubing connector into the socket in the left side of the monitor Make certain that the connector is completely inserted to avoid air leakage duri...

Page 7: ...rtably Measurement a Be seated with your feet flat on the floor and don t cross your legs b Place palm upside in front of you on a flat surface such as a desk or table c The middle of the cuff should...

Page 8: ...f the monitor manually h During measurement you can press the exhaust valve which is located in the front of the bulb to release the air Note Please consult a health care professional for interpretati...

Page 9: ...itor 8 TROUBLESHOOTING 2 PROBLEM POSSIBLE CAUSE SOLUTION LCD Display shows EE The cuff was not properly applied or the rubber tube was bent or pressed Review the cuff applying and testing sections of...

Page 10: ...iately qualified technical personnel to repair those parts of equipment which are designated repairably can be supplied 9 The monitor can maintain the safety and performance characteristics for a mini...

Page 11: ...use in the electromagnetic environment specified below The customer or the user of the KD 322 should assure that it is used in such an environment Emissions test Compliance Electromagnetic environmen...

Page 12: ...equency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Table 3 For ME EQUIPMENT and ME SYSTEMS that are not LIFE SUPPORTING Gu...

Page 13: ...radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic enviro...

Page 14: ...12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be de...

Reviews: