English 6
Symbols that appear on the display
Symbols
Function / Meaning / Recommended Action
Appears while measurement is in progress. It blinks when the pulse is detected.
Remain as still as possible.
IHB/AFib symbol appears when an irregular heartbeat is detected.
It may light when a very slight vibration like shivering or shaking is detected.
Symbols that appear on the display (continued)
Symbols
Function / Meaning
Recommended Action
Previous measurements stored in MEMORY
Average data
FULL BATTERY
The battery power indicator during measurement
LOW BATTERY
The battery is low when it blinks
Replace all batteries with new ones
when the indicator blinks.
or
Unstable blood pressure due to movement during measurement
Take another measurement.
Remain still during measurement.
The systolic and diastolic values are within 10 mmHg of each other.
Apply the cuff correctly, and take
another measurement.
The pressure value did not increase during inflation.
The cuff is not applied correctly.
PUL.DISPLAY ERROR
The pulse is not detected correctly.
Blood pressure monitor internal error
Remove the batteries and press the
START
button, and then install the
batteries again. If the error still
appears, contact the dealer.