![AND UA-705 Instruction Manual Download Page 14](http://html.mh-extra.com/html/and/ua-705/ua-705_instruction-manual_2941361014.webp)
English 13
Technical Data
Type UA-705
Measurement method
Oscillometric measurement
Measurement range
Pressure:
20 - 280 mmHg
Pulse:
40 - 200 beats / minute
Measurement accuracy Pressure:
±3
mmHg
Pulse:
±5%
Power supply
1 x 1.5V batteries (R6P,
LR6 or AA)
Number of measurements
Approximately 4000 measurements, when AA
Alkaline batteries are used, with pressure value of
180 mmHg at room temperature of 23 °C.
Classification
Internally powered ME equipment
Continuous operation mode
Clinical test
According to ANSI / AAMI SP-10 1987
EMC IEC
60601-1-2:
2007
Data memory
Last 30 measurements
Operating conditions
+
10 °C to
+
40°C / 15 %RH to 85 %RH
800 hPa to 1060 hPa
Transport / Storage conditions
-
10 °C to
+
60 °C / 15 %RH to 85 %RH
Dimensions Approx.
81
[W] x 54 [H] x 105 [D] mm
Weight
Approx. 120 g, excluding battery
Applied part
Cuff Type BF
Useful life
Device: 5 years (when used six times a day)
Cuff:
2 years (when used six times a day)
Accessories sold separately
Arm
size:
The
circumference at the biceps.
Note: Specifications
are subject to change for improvement without prior notice.
EMC table information is listed on our website:
http://www.aandd.jp/products/manual/medical/emc_en.pdf
Cuff
Catalog Number
Cuff Size
Arm Size
CUF-D-LA
Large adult cuff
32 cm to 45 cm
CUF-D-A
Adult cuff
22 cm to 32 cm