E-22
MAINTENANCE
Do not open the device. It uses delicate electrical components and an
intricate air unit that could be damaged. If you cannot fix the problem
using the troubleshooting instructions, contact the authorized dealer in
your area or our customer service department. A&D customer service
can provide technical assistance and spare parts.
Type UA-651BLE-V Series
Measurement method Oscillometric measurement
Measurement range Pressure: 20 - 280 mmHg
Systolic pressure: 60 - 279 mmHg
Diastolic pressure: 40 - 200 mmHg
Pulse: 40 - 180 beats /min
Measurement accuracy Pressure: ±3 mmHg
Pulse: ±5 %
Power supply 4 x 1.5V batteries (R6P, LR6 or AA)
or AC adapter (TB-233) (Not included)
Number of measurements Approx. 700 times LR6 (alkaline batteries)
Approx. 200 times R6P (manganese batteries)
With pressure value 180 mmHg, room
temperature 23 °C.
Classification Internally powered ME equipment (Supplied
by batteries) / Class II (Supplied by adapter)
Continuous operation mode
Clinical test According to ANSI / AAMI SP-10 2002
Wireless communication VZ (MURATA Manufacturing Co.Ltd.)
Bluetooth Ver. 4.0LE BLP
TECHNICAL DATA