Problem
Reason
How to fix
Amount dispensed is
too much
Liquid on outer part of
tip
Refer to “10 PIPETTING FOR
ACCURATE DISPENSING”
Amount dispensed is
too little
Solvent evaporates
and increases
pressure inside the tip.
Refer to “10 PIPETTING FOR
ACCURATE DISPENSING”
Device produces an
abnormal noise
Piston becomes stuck
When the pipette
hasn’t been used for a
while inner piston parts
can become stuck due
to grease.
After moderate usage device
should return to normal
functionality.
Tip won’t eject
Tip length is incorrect
Refer to “12 ADJUSTING
TIP EJECTOR HEIGT”
Tip holder discoloration Dispensing acids for
long times.
If device functionality is affected
replace the lower part.
01 Err
Connection nut (Refer
to “7-5 Parts names
and materials”)
is loose
Reattach the connection nut.
Press the reset key and reset
the device.
02 Err
98 Err
Stepping motor failure
Press the reset key and return
the motor position to the origin