20
When carrying out pre-rinse for multiple dispensing or stopping multiple dispensings,
use the total discharge function. (Refer to “9-8. Total discharge function”)
Minimum dispensing amount and maximum dispensing count for multiple dispensing
mode is as follow.
MODEL
Minimum dispensing amount
Maximum dispensing count
MPA-10 0.3
μ
L 33
times
MPA-20 0.3
μ
L 66
times
MPA-200 3
μ
L 66
times
MPA-1200 15
μ
L 80
times
MPA-10000 0.1mL
99
times
MPB-200-8 3
μ
L 66
times
MPB-1200-8 15
μ
L 80
times
2) Selecting the multiple dispensing mode
(The display is an example.)
[ 1 ] Press the
Back key.
[ 2 ] Press the
Up or Down key to select ”MD”.
[ 3 ] Press the Enter key to select multiple
dispensing mode (MD).
[ 4 ] Press the Up or Down key to set dispensing
amount for one time.
[ 5 ] Press the Enter key to confirm dispensing
amount for one time.
[ 6 ] Press the Up or Down key to set dispensing
count.
* “Dispensing amount for one time x dispensing
count” can not be set if it exceeds the volume range.