background image

Sub MkIIP Oxygen & Carbon Dioxide Monitor 
User Manual - US Navy Version 

 

Document Ref

: AS2-802-14 - 

January 2016 

Page 14 

  

 

Disable Logging: 

If  you  do  not  want  data  logging  to  take  place  at  all,  press  the  ‘Disable  Logging’  button. 
‘Logging disabled’ will be shown on the Instrument Status page.  
 

Logging Options: 

Select  the  data  items  that  you  wish  to  record  from  O

2

,  CO

2

,  Depth  and  Temperature,  by 

clicking  on  the  box  next  to  the  parameter  you  require,  if  the  box  has  been  successfully 
selected a 

 will appear.   

 

Logging Resolution: 

Select  the  Logging  Resolution  you  require  by  clicking  on  the 

▼  arrow, and then clicking on 

the  time  interval  you  want.  This  is  the  time  interval  between  logged  readings,  which  can  be 
set to one of various values between 10 seconds and 30 minutes.  Please be aware that the 
shorter the time interval, the shorter the data storage period. For example logging just CO

2

 at 

2 minute intervals will give approximately 4 days of continuous data storage, where as logging 
O

2

, CO

2

, Depth and Temperature at 10 second intervals will result in approximately 2 hours of 

continuous  dat

a  storage.  The  wording  ‘continuous’  refers  to  the  unit  logging  when  left 

permanently switched on. If the instrument is switched off, e.g. overnight, this time period will 
be extended.  
 

Note  there  is  a  slight  overhead  every  time  the  instrument  is  switched  on  and  off,  so  the 
number of data points recorded will be slightly reduced each time the instrument is switched 
off and on.  In normal daily use this reduction will not be apparent. 

 

Full Memory: 

This  option  enables  you  to  specify  what  the  instrument  should  do  when  its  data  logging 
memory is full. 
 
You may choose to ‘Ignore newer data’ if for example you are conducting a specific test run, 
using  a  short  logging  resolution,  and  you  only  require  data  from  that  test.  By  ignoring  new 
data you will ensure that you retain your test data until you have been able to download it to a 
PC.  
 
Alternatively you may choose to ‘Discard older data’ if you want the instrument to essentially 
keep on logging for ever, and you only intend to download the data following an incident you 
want to investigate.  
 
Once you have set each of the 3 options described above (Logging Data, Logging Resolution 
and Memory Full options), press ‘Clear Instrument and Setup for Logging’. This transfers the 
selected  options  to  the  instrument.    It  also  transfers  the  date  and  time  from  your  PC  to  the 
instrument.  Make sure the Date and Time on your PC are correct! 
 
You will be asked to confirm whether logging should ‘Start Immediately’, or the next time the 
instrument  is  switched  on  (‘Start  at  next  Power-Up’).  This  enables  you  to  optimize  the  data 
storage time by leaving the instrument switched off until it is required.  
 

Reset log with Elapsed Time: 

Select this option if you wish to enable the Reset log with Elapsed Time option, by clicking on 
the box. If the box has been successfully selected a 

 will appear. 

 
With this option enabled the data log period will be reset and any existing data in the data log 
memory  cleared,  when  the  Elapsed  Timer  is  reset  to  zero  (see  Section 

Error!  Reference 

source not found.

).  Enabling this option will ensure that data is logged during any DISSUB 

incident for later analysis. 
 
 
 
 

Summary of Contents for SA1 W75 SR 60 N 564

Page 1: ...ck Court Stokesley Business Park North Yorkshire TS9 5PT UK T 44 0 1642 711400 F 44 0 1642 713900 W www analox net E info analox net Sub MkIIP Oxygen Carbon Dioxide Monitor User Manual US Navy Version...

Page 2: ......

Page 3: ...al power 100 240VAC 8 6 Calibration 9 6 1 Readout check 9 6 2 Use of calibration gas 10 7 Data logging 11 7 1 Installation of data logging software on a PC 11 7 2 Data logging program overview 13 7 3...

Page 4: ...Sub MkIIP Oxygen Carbon Dioxide Monitor User Manual US Navy Version Document Ref AS2 802 14 January 2016 Page 2...

Page 5: ...supply cable Part Number SAEP AS2 605 c User manual d Test certificates e Analox hyperbaric tested batteries x4 Part Number 2557 4625 Optional f Data output cable Part Number SAEP AS2 602 and softwar...

Page 6: ...ch the standard instrument has only one operator control A single pushbutton controls backlights for each of the liquid crystal displays LCD and can also be used to reset the elapsed time The unit is...

Page 7: ...the fuses in the external supply line or the battery circuit are not blown On powering up the instrument the sensor readings may take a short time to settle particularly the CO2 sensor which can take...

Page 8: ...nstrument is operating from an external power supply The backlight is turned on by momentarily pressing the pushbutton on the front of the instrument The backlight will turn off automatically after a...

Page 9: ...nt damage open the lid of the instrument using the two catches beside the handle b Undo the two Velcro retaining straps around the batteries c Ease each battery from its clips use one hand to grip the...

Page 10: ...ypical current consumption figures are shown below Input Voltage V DC Load Current without backlight mA Load Current with backlight mA 9 0 160 315 12 0 105 340 24 0 55 290 40 0 37 270 5 2 External pow...

Page 11: ...the right hand column as a guideline to the acceptable tolerance Also ensure that the SUB MkIIP is sampling the atmosphere from the same compartment as the CAMS To convert the CAMS Gas reading from u...

Page 12: ...ed with the instrument will be required and should be fitted as shown in Figure 2 The longer free pipe should be connected to the outlet of a test gas bottle flow regulator The flow adaptors in the ch...

Page 13: ...ensure all data from a DISSUB event is captured A cable is supplied to connect the SUB MkIIP to the serial port on a PC The cable is fitted with a 9 way female D type connector An adaptor will also b...

Page 14: ...ED THE NEXT TIME YOU START THE PROGRAM 11 Always ensure that the Baud rate is set to 9600 to which it will default 12 Click on the Communicate with Instrument button You will see a message Please wait...

Page 15: ...iable that you wish to view and a graph will be automatically generated This process will suspend logging you will therefore need to click on the Restart logging button on the Instrument Status page T...

Page 16: ...be apparent Full Memory This option enables you to specify what the instrument should do when its data logging memory is full You may choose to Ignore newer data if for example you are conducting a s...

Page 17: ...spreadsheet Therefore select an Excel compatible csv file CSV stands for a Comma Separated Variable file and is a format that most spreadsheet programs will recognize 8 Now select the folder and enter...

Page 18: ...n the standard width of the cell Click on the A at the top of the column and then select Format Column and Autofit Selection Or simply drag the column separator between A and B to achieve the right wi...

Page 19: ...calibration is required at least once every 12 months to replace the electrochemical oxygen sensor Where the instrument is used as a standby for use in emergencies replace the main batteries as detail...

Page 20: ...effects Body Part Effect First Aid Procedures Skin Contact could result in a chemical burn Wash the affected part with a lot of water and remove contaminated clothing If stinging persists get medical...

Page 21: ...cklight times out after 15 seconds when turned on Oxygen Sensors See Appendix on page 22 Analox 9100 9212 9HSUB oxygen sensor with microprocessor applied temperature compensation 2 3 year life at 21 S...

Page 22: ...replacement or a refund of net purchase price paid as per our Invoice on initial purchase from us We shall have no liability for losses damages costs or delays whatsoever We shall have no liability f...

Page 23: ...Version Document Ref AS2 802 14 January 2016 Page 21 12 Disposal According to WEEE regulation this electronic product can not be placed in household waste bins Please check local regulations for info...

Page 24: ...yields the same partial pressure of oxygen as that present in the current environment SEV on the Sub MkIIP uses 1000mbara as its surface pressure for a point of reference For example 1 If the atmosph...

Page 25: ......

Reviews: