Carefully pull the electrical connector from the rear of the Carbon
Monoxide sensor.
Fit the new Carbon Monoxide sensor to the connector, note it will only
connect in one orientation.
NOTE:
THE NEW CARBON MONOXIDE SENSOR IS SUPPLIED IN
A HOUSING, READY TO BE FITTED.
Carefully feed the sensor wire and sensor housing through into the
bulkhead and then firmly push the housing into the Bulkhead on the
under-side of the Analox CO
Clear
enclosure.
Switch the instrument back on.
Allow the sensor to settle for 1 hour.
Perform a Zero and Span Calibration Check to calibrate the new sensor.
Test the operation of the Carbon Monoxide alarms (see Alarm Check
section).