background image

Evaluation Board User Guide 

UG-015

 

Rev. 0 | Page 7 of 12 

REAL-TIME MEASUREMENT TAB

 

The 

Real-Time Measurement

 tab configures the inertial sensor 

evaluation system and the ADXL345 for real-time acceleration 
monitoring. The tab contains an oscilloscope-like interface that 
you can use to view the output of the accelerometer and adjust 
the relevant parameters, such as bandwidth, range, and offset  
(see Figure 12).  

0

81

19

-0

09

 

Figure 12. Real-Time Measurement Tab 

Within the 

Real-Time Measurement

 tab, additional tabs exist 

for displaying the output in terms of least significant bits (LSBs) 
or acceleration in 

g

’s. In the 

Output – LSB

 tab, the output range 

can be adjusted by using the graph tools in the upper right 
corner or by selecting the x-axis (labeled 

Sample

)

 

or y-axis 

(labeled 

Output [LSB]

) limits and typing a new value. For 

example, selecting 1024 on the y-axis and typing 2048 adjusts 
the scale to show +2048 LSBs to −1024 LSBs. 

The configuration section is at the bottom of the 

Real-Time 

Measurement

 tab. In the 

Data Rate and Resolution

 box, the 

output data rate and 

g

 range can be adjusted. Selecting the  

Full-Resolution

 checkbox places the part into full-resolution 

mode where the scale factor is fixed. Deselecting this checkbox 
places the part into fixed 10-bit mode. For more information 
about the operation of the part, refer to the ADXL345 data sheet.  

The 

Self Test

 button toggles the self-test bit of the ADXL345. 

When self-test is activated, the sensor beam is deflected. The 
electronics detect this by means of a shift in all three axes. When 
self-test is enabled, the indicator next to the 

Self Test

 button is 

lit. Clicking the 

Self Test

 button when self-test is enabled turns 

the self-test feature off and causes the 

Self Test

 indicator to dim. 

Self-test is enabled as long as the indicator is lit. 

The ADXL345 provides three offset registers that are used  
for calibration and offset adjustment. On the 

Real-Time 

Measurement

 tab, these registers are easily accessible for 

programming offset values. This can be done in the 

Offset 

Adjustment

 box by typing a value into the text box below the 

corresponding axis knob or by clicking and holding the knob  
and rotating it on the screen with the mouse cursor. The value 
entered is directly written into the offset registers. 

After configuration, you can begin real-time measurement by 
clicking the 

Start Meas

 button. This causes many of the options 

and tabs to be grayed out or to disappear, to prevent software 
conflicts, until the 

Stop Meas

 button is clicked. The accelerometer 

output data then begins to flow across the screen at the selected 
output data rate. 

Up to the last 8192 samples (x-, y-, and z-axis) of accelerometer 
output data can be saved to a .txt file. Each file created contains 
a header with the date, bias values, and accelerometer data, in 
LSBs, aligned in tab-delimited columns. To save the data, you 
must enable the 

Save Data

 button. Click this button and type  

in or select a file name with a .txt extension. 

Each time a measurement is performed, the new data is appended 
to the old data. This means that saving data also saves old data, 
if new data is less than 8192 samples, unless the old data is first 
cleared. The data saved to a file is the LSB data; therefore, the 

Output – LSB

 window must be cleared so that old data is not 

saved. To clear the 

Output – LSB

 window, right-click the x-axis 

(labeled 

Sample

) in the space between 

Sample

 and the window 

and select the 

Clear Chart

 option. 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for iMEMS ADXL345

Page 1: ...s directly to a PC via its USB where the USB connection provides both power and communications to the board The ISEB then connects to the satellite board through a ribbon cable This cable allows the satellite to be easily manipulated for testing or for being separately placed into an environmental chamber for temperature or humidity testing Separating the boards mitigates corruption of data due to...

Page 2: ...Setup 3 Software Configuration 3 Hardware Configuration 4 Updating Repairing the ISEB Firmware 5 Inertial Sensor Evaluation System Graphical User Interface 6 Configuration Tab 6 Real Time Measurement Tab 7 Interrupt Configuration Tab 8 Power Consumption Tab 9 Temperature Tab 10 Header Pinout 11 ESD Caution 11 REVISION HISTORY 6 09 Revision 0 Initial Revision ...

Page 3: ...see Figure 2 08119 002 Figure 2 Found New Hardware Prompt 3 Select Don t search I will choose the driver to install see Figure 3 and select ADI Inertial Sensor Evaluation System from the model list see Figure 4 The ISEB should be detected automatically in the Device Manager as the ADI Inertial Sensor Evaluation System under the Ports COM LPT selection It is recommended to open the Device Manager t...

Page 4: ...e ribbon cable 5 Verify that the jumpers located in the center of the ADXL345 satellite are configured as shown in Figure 5 CONNECTOR ADXL345 JUMPER POSITIONS 08119 005 Figure 5 Correct Jumper Positions for the ADXL345 Satellite 6 Place the ADXL345 ultralow power accelerometer into the socket on the satellite board The Socket Pin 1 Indicator is located inside the socket close to the hinge as shown...

Page 5: ...hen the COM port is selected click OK to accept the changes and go back to the main screen 08119 108 Figure 8 Selecting the Correct COM Port for the Downloader When the ISEB is connected the correct firmware is selected and the downloader is fully configured follow these steps to flash the firmware 1 Click Start as shown in Figure 7 to initiate the flashing process After clicking Start the two but...

Page 6: ...I Inertial Sensor Evaluation System under the Ports COM LPT selection 08119 007 Figure 10 Selecting the Correct COM Port for the ISEB After selecting the COM port click on Connect to connect to the ISEB This causes the indicator under Vs Set to light up and update the Current Vs V value verifying that the board is connected see Figure 11 08119 008 Figure 11 ISEB Connected to PC Clicking the round ...

Page 7: ...licking the Self Test button when self test is enabled turns the self test feature off and causes the Self Test indicator to dim Self test is enabled as long as the indicator is lit The ADXL345 provides three offset registers that are used for calibration and offset adjustment On the Real Time Measurement tab these registers are easily accessible for programming offset values This can be done in t...

Page 8: ...ils refer to the ADXL345 data sheet Clicking an indicator causes it to toggle values A lit indicator causes a value of 1 to be written to that bit and an unlit indicator causes a value of 0 to be written to that bit In the Interrupt Configuration box you can select the interrupts to enable For purposes of the GUI all enabled interrupts are mapped to the INT2 pin Selecting the checkbox next to the ...

Page 9: ...n indicator lights up signifying that a controller or microprocessor is actively pulling data from the part or processing it In addition to power savings from the ADXL345 more power can be saved by powering down the microprocessor during periods of inactivity and using the ADXL345 activity interrupt to wake the microprocessor up during periods of activity The ACT INACT_CTL indicators can be used t...

Page 10: ... Figure 15 allows a maximum number of points to be set and at what rate data is acquired The testing can be stopped before reaching the maximum number of points The data rate for the accelerometer can also be set to ensure that the testing is performed under conditions most like those that the part would be configured for in a system Figure 15 Temperature Tab The data in the top graph in Figure 15...

Page 11: ... HEADER PINOUT 1 S1 2 3 S2 4 5 S3 6 7 S4 8 CS2 9 VS2 10 VS1 11 12 ST1 13 ST2 14 SCL1 15 SDA1 16 CLK 17 MOSI 18 MISO 19 CS3 20 CS4 P1 3 3V CS2 VS2 VS1 ST1 ST2 SCL1 SDA1 CLK MOSI MISO CS3 CS4 AGND GND 08119 013 Figure 16 ISEB 20 Pin Header Pinout ESD CAUTION ...

Page 12: ... use of evaluation boards Information furnished by Analog Devices is believed to be accurate and reliable However no responsibility is assumed by Analog Devices for its use nor for any infringements of patents or other rights of third parties that may result from its use Analog Devices reserves the right to change devices or specifications at any time without notice Trademarks and registered trade...

Reviews: