UG-595
Evaluation Board User Guide
Rev. 0 | Page 6 of 12
EVALUATION BOARD OVERVIEW
Figure 5.
LFCSP Demo Board Typical Operation Setup
TYPICAL OPERATION
The typical test setup for the
charger consists of a dc
power supply unit (PSU) for VIN_F, a source meter unit (SMU)
or a battery simulator for the ISO_Bx pins, and a variable power
resistor or electronic load for the ISO_Sx pins.
The SMU at the ISO_B_F node must have a 100 mΩ to 250 mΩ
resistor (R
S
) in series with its positive lead. The resistor emulates
the equivalent series resistance of a real battery. Some SMU
models that have been successfully used for the ISO_x_F node
include the following:
•
Keithley 2306 battery simulator
•
Keithley 2602A SMU
•
Agilent 6784A/6762A SMU
R3
C5
C4
JP1
J3
J11
U2
JP2
JP3
C3
D1
J1
1
J2
1
9
8
7
2
19
18
17
16
15
14
13
12
11
10
1
TP4
TP3
TP6
TP5
TP20
C12
C6
U1
U3
J8
J10
J7
J6
J5
R18
R17
R16
R6
R13
R5
R2
R8
R15
R14 R1
R7
R11
R10
THR
5
20
16
14
20
10
CBP
SYS_EN
BAT_SNS
VCO_LDO
ISO_B_S
ISO_S_S
D3
D1
D2
IL
ED
ISO
_
S
_
F
V
DDI
O
D3
D2
D1
GND
BAT_SNS
THR
SYS_EN
GND
VIN_F
GND
ILED
GND
GND
GND
D
3
S
D
3
D1S
VIN_F
ISO_S_F
ISO_B_F GND
GND
SCL
GND
GND
ISO_S_F
ISO_S_F
ISO_S_S
ISO_B_S
ISO_B_S
VIN_S
GND_S
THR
ILED_S
AGND
BAT_SNS
ISO_B_F
SDA
D2S
D2
D1
VIN_S
SDA
VIN_F
SCL
1
1737-
005
SMU
O
R
BAT
T
E
RY
G
ND
BAT
T
E
RY
TH
E
R
M
IS
TOR
R
S
SYST
EM L
O
A
D
G
ND
V
IN 5.
0V
G
ND
C1
C2
ADP
5063 Q
F
N20 DE
M
O