Adjust flow rate

- - -

 

mL/h

Value must be larger than zero

11:09

PM

Adjust flow rate

125

 

mL/h

Next 

11:09

PM

Select volume option

Enter

volume

Use

entire bag

11:09

PM

Adjust volume

500

 

mL

Next 

11:09

PM

1

Press   and   to adjust flow rate from 1 to 400 mL/h, then press the multifunction button [ ].

Press   to use entire bag. Press   to set the volume to be delivered. This value must be from 1 to 
9,999 mL. Then press the multifunction button [ ].

Review the feeding parameters (with or without a programmed volume). Then press the 
multifunction button [ ].

Press the multifunction button [ ] to start the feeding.

<<Simplified>> Mode

2

The following parameters may be modified 
when the pause mode is activated in 
simplified mode: 

 The flow rate

 The volume programmed

Press and hold the multifunction button  
[ ] key for 3 seconds and the progress bar 
fills up (green bar at the bottom of  
the screen).

Press the   button to access the flow  
rate screen. 

<<Pause>> Mode

Using the Pump

Running

125

 

mL/h

Hold to pause

11:09

PM

Running

125

 

mL/h

500

 

mL

Hold to pause

0

 

mL

h

 

00 

min

Rate

Time left

Total

Total Fed

11:09

PM

Change rate or volume

Resume nutrition

Paused

0

 mL

125

 

mL/h

500

 

mL

h

 00 

min

Rate

Total

Total Fed

11:09

PM

Time left

Summary of Contents for PUGGLE

Page 1: ...Quick Reference Guide Enteral Feeding Pump ...

Page 2: ... and volume to be administered 4 Lower part of the screen Displays time left and volume administered 5 Action option To pause the feeding press and hold the multifunction button until the green progress bar fills entirely Running 125mL h 1000mL Hold to pause 450mL 4 h 24 min Rate Time left Total Total Fed 11 09PM 1 2 3 4 5 Icons Explanation The flow rate In mL h from 1 to 400 mL h The volume progr...

Page 3: ...ubes down in line with their respective slots Close the pump cover then place the adapter into the designated slot Remove the protective cap from the distal connector end of downstream tube that goes on the patient s feeding tube Launch automatic priming see step 3 Connect the tubing to the patient enteral feeding tube and program the pump Installation of the Tubing Turn on the PUGGLE pump by hold...

Page 4: ...he tubing Switch on the pump by pressing and prime tubing by pressing Disconnect patient then press to prime Check tubing is correctly primed Restart prime by pressing if needed or program nutrition by pressing then connect the tubing to the patient Prime tubing before starting nutrition Program nutrition Prime 11 09PM Disconnect Patient Prime 11 09PM Restart prime Prime stopped Check tubing is co...

Page 5: ...lume Then press the multifunction button Press the multifunction button to start the feeding Simplified Mode 2 The following parameters may be modified when the pause mode is activated in simplified mode The flow rate The volume programmed Press and hold the multifunction button key for 3 seconds and the progress bar fills up green bar at the bottom of the screen Press the button to access the flo...

Page 6: ...e 11 09PM Charge the battery by connecting the AC adapter to the pump and power source Running Low battery Less than 30 min remaining Dismiss notice 11 09PM Wait until the end of the feeding or Pause the pump to modify the feeding settings or Switch the pump off Solid yellow LED Low priority alert ...

Page 7: ...he multifunction button to program a new feeding Relaunch pause or resume feeding In case of an unexpected Air alarm check that the tubing is properly positioned in its slot disconnect from patient then launch another automatic prime Flashing yellow LED Medium priority alarm ...

Page 8: ... occlusion Make sure the tubing is positioned correctly If there is an occlusion between the pump and the patient release the occlusion and ensure that the branch of the cassette is empty If alarm is not resolved check that the cassette is positioned correctly between the two black pins and that there is no presence of feeding solution between the two black pins If the alarm is still not resolved ...

Page 9: ...eshooting Issue Solution Contact customer service or your healthcare provider Contact customer service or your healthcare provider Contact customer service or your healthcare provider Flashing red LED High priority alarm ...

Page 10: ...hooting Issue Solution Power indicator light remains off Request a new AC adapter Issue Solution Continuous beeping sound Contact customer service or your healthcare provider Flashing red LED High priority alarm Other alarms ...

Page 11: ...ch crevices use a soft brush It is recommended that the pump be cleaned to eliminate stains and soils after each tubing use or at least every day A cleaning is also necessary when the pump is allocated to a new patient Battery Charging time 2 5 hours Autonomy 20h at 125 ml h Memory Specific values and modes 24h For further information please refer to the instructions for use Contact Amsino s custo...

Page 12: ...g of the indications contraindications warnings and precautions The symbol indicates the trademark is registered in the U S Patent and Trademark Office and certain other countries Copyright 2021 Amsino International Inc All rights reserved Amsino International Inc 708 Corporate Center Drive Pomona CA 91768 Amsino com Phone 800 632 6746 Fax 909 626 3888 Email info amsino com ...

Reviews: