5-15
Troubleshooting
Routine Maintenance
129367-410
Alarm
Description
Screen with Operator Instructions
Reference Drain
Deviation Alarm
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM! REFERENCE
249
DRAIN DEVIATION ALARM
CHAMBER:
000.0 C
00.0 psig
REFER:
000.0 C
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
Occurs if chamber
temperature varies
±2°F (1°C) from the
reference temperature.
(Lab/Sci program option
only.)
Screen with Service Instructions
Screen with Operator Instructions
→
DEVIATION BETWEEN REFERENCE AND
CHAMBER PROBE IS MORE THAN 1 C.
CAUSES AND CORRECTION:
1.
LOOSE CONNECTION IN PROBE WIRING
→
REPAIR
2.
CHAMBER/REFERENCE PROBE FAILED
→
REPLACE
→
RECALIBRATE
3.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4.
MAIN CONTROL FAILURE
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
REPLACE 146656-185 CONTROL BOARD
→
REPLACE 146659-003 ANALOG BOARD
→
RECALIBRATE
STATUS..
SERVICE INFORMATION:
250
REFERENCE/CHAMBER DEVIATION
PAPER
CLEAR
EXIT
FEED
ALARM
Waste Temperature
Probe Failure
Occurs if waste line
temperature reading is
outside the normal range
of 32-293°F (0-145°C).
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM! WASTE
215
TEMPERATURE PROBE FAILURE
CHAMBER:
000.0 C
00.0 psig
WASTE:
000.0 C
STERILIZER WILL:
•
PREVENT NEW CYCLE FROM BEING
STARTED UNTIL ALARM IS CLEARED
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE