5-5
Troubleshooting
Routine Maintenance
129367-410
Alarm
Description
Screen with Operator Instructions
Occurs if chamber does
not reach the set
evacuation level within
the allotted time.
Too Long In Evacuation
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK WATER SUPPLY VALVE
→
IF CLOSED, OPEN VALVE
3. IF ALARM RECURS, ABORT CYCLE
AND CALL SERVICE
STATUS..
ALARM!
202
TOO LONG IN EVACUATION
For Healthcare
Only
STATUS
PAPER
FEED
SILENCE
ABORT
SERVICE
ALARM
HELP
For Life Sciences
Only
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK WATER SUPPLY VALVE
→
IF CLOSED, OPEN VALVE
3. IF ALARM RECURS, ABORT CYCLE
AND CALL SERVICE
STATUS..
ALARM!
202
TOO LONG IN EVACUATION
SILENCE
ABORT
SERVICE
ALARM
HELP
ABORT
STATUS
PAPER
MORE
FEED
HELP
STATUS..
SERVICE INFORMATION:
203
TOO LONG IN EVACUATION
→
CHAMBER DID NOT REACH REQUIRED
VACUUM LEVEL WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
1.
WATER PRESSURE LESS THAN 30 PSIG
→
CHECK WATER SUPPLY PIPING
2.
CHAMBER DRAIN STRAINER PLUGGED
→
CLEAN
3.
CHECK VALVE MALFUNCTION
→
REPAIR
4.
SOLENOID VALVE MALFUNCTION
→
REPAIR
5.
DOOR SEAL NOT ACTIVATED
→
CHECK SEAL
→
CHECK SEAL STEAM AND EXHAUST
NOTE: This alarm has
two service help screens.
For Life Sciences
Only
Screen with Service Instructions