6) Press the RED power button to start the heating phase of the cycle.  
 
7) The LED will flash RED for the duration of the heating phase. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
8) KEEP ALL PERSONNEL AWAY FROM THE FRONT OF THE COOL-X. 
  

CAUTION:  X-RAYS ARE PRESENT 

 

9) When the cooling phase starts, the LED will automatically start flashing GREEN.  This 
pattern of flashing red/green will continue until the red power button is pressed again.  Power to 
the unit is toggled ON/OFF every time the power button is pressed.  When the power is OFF, the 
LED is also OFF. 
 
10) At the start of a cycle and in between the heating/cooling phases a “BEEP” sound will be 
heard from a small internal speaker to indicate the transition of phases.  If the sound interferes 
with the x-ray detector due to microphonic pick-up, the speaker can be disconnected.  To 
disconnect the speaker, remove the cover by removing the 4 small screws from the top.  Locate 
the small black jumper marked “Beep”.  Move the jumper over one pin to eliminate the sound.  
 
 
 
 
 
 
 
 
 
 
 

 

 

Heating phase 

Cooling phase 

 

Area of enlargement 

Speaker Jumper 

Figure 4 

Figure 5 

Figure 6 

Summary of Contents for COOL-X

Page 1: ...COOL X X Ray Generator with Pyroelectric Crystal Operating Manual AMPTEK INC 14 DeAngelo Drive Bedford MA 01730 2204 USA 1 781 275 2242 Fax 1 781 275 3470 Amptek sales ametek com www amptek com ...

Page 2: ...tion Safety for X Ray Diffraction and Fluorescence Analysis Equipment for more information see http www fda gov cdrh radhlth pdf xdfgehb pdf Beryllium Window Caution The front of the COOL X is made of a 10 mils 250 µm Beryllium window DO NOT REMOVE the metal cover from the COOL X until ready to use The Be window is extremely brittle and can shatter easily Do not have any object come into contact w...

Page 3: ... defects or other reproductive harm if exposed to them through improper use storage or disposal of the product Prop 65 Chemical Type of Toxicity CAS No Product part containing the chemical Beryllium Cancer Detector window Please consult this owner s manual for proper use storage care and disposal of the product For more information go to www p65warnings ca gov ...

Page 4: ...nerator is completely enclosed inside the small TO 8 package at the tip of the extender TURN OFF POWER TO THE COOL X WHEN NOT IN USE 2 The Controller is enclosed inside the metal box The function of the Controller is to supply power to the X ray generator monitor its temperature and automatically change the heating and cooling phases of the cycle Figure 1 Please see Appendix A if the COOL X emitte...

Page 5: ... is with the AC adapter If battery operation is required remove the 4 small screws from the top of the box and insert the provided 9 Volt battery to the terminals 4 The COOL X should be mounted on a firm surface using the 4 mounting tabs of the box 5 Connect the AC adapter to the back of the box and remove the front metal cover Power On Off AC Adapter Jack input Flashing LED Figure 2 Figure 3 ...

Page 6: ...unit is toggled ON OFF every time the power button is pressed When the power is OFF the LED is also OFF 10 At the start of a cycle and in between the heating cooling phases a BEEP sound will be heard from a small internal speaker to indicate the transition of phases If the sound interferes with the x ray detector due to microphonic pick up the speaker can be disconnected To disconnect the speaker ...

Page 7: ...tory This warranty does not apply in the event of misuse or abuse of the instrument or unauthorized alterations or repair AMPTEK INC shall not be liable for any consequential damages including without limitation damages resulting from the loss of use due to failure of this instrument The factory MUST be notified prior to return shipment All products returned under the warranty must be shipped prep...

Page 8: ... pattern for the TO 8 for attaching it to the heat sink is provided below The Control electronics should also be mounted on a metal plate and kept at approximately the same temperature as the TO 8 This is important because the temperature sensor in the Control circuit refers to a matched temperature sensor inside the TO 8 emitter head in order to maintain the proper temperature differentials and t...